background image

Type

Battery

Voltage

Capacity

GEB222

Li-Ion

7.4 V

6.0 Ah

Type

Battery

Voltage

Capacity

GEB371

Li-Ion

13 V

16.8 Ah

Temperature

Type

Operating temperature

[°C]

Storage temperature

[°C]

All instruments

-20 to +50

-40 to +70

Leica SD cards

-40 to +80

-40 to +80

Battery internal

-20 to +55

-40 to +70

Protection against water, dust and sand

Type

Protection

All instruments

IP55 (IEC 60529)

Humidity

Type

Protection

All instruments

Max 95 % non condensing

The effects of condensation are to be effectively

counteracted by periodically drying out the instru-

ment.

Type

Additive

Constant [mm]

ATR

SpeedSearch

Standard prism,

GPR1

0.0

yes

yes

Mini prism,

GMP101

+17.5

yes

yes

360° prism,

GRZ4 / GRZ122

+23.1

yes

yes

360° Mini prism,

GRZ101

+30.0

yes

not

recommended

Reflector tape

S, M, L

+34.4

yes

no

Reflectorless

+34.4

no

no

There are no special prisms required for ATR or for SpeedSearch.

Working range:

5 m to 150 m (15 ft to 500 ft)

Position accuracy:

5 cm at 100 m (1.97" at 330 ft)

Internal battery

External battery

Environmental

specifications

Reflectors

Electronic Guide Light

EGL

Technical Data

63

Содержание iCON iCR70

Страница 1: ...Leica iCON iCR70 User Manual Version 1 0 English...

Страница 2: ...cal data and safety directions Intended as a quick reference guide iCR70 User Manual All instructions required in order to operate the product to a basic level are contained in the User Manual Provide...

Страница 3: ...ownload your latest calibration cer tificates and service reports mySupport Create new support requests for your products that will be answered by your local Leica Geosystems Support Team View the com...

Страница 4: ...Data Storage Concept 26 2 3 Container Contents 27 2 4 Instrument Components 28 3 User Interface 30 3 1 Keyboard 30 3 2 CommunicationHandle 32 4 Operation 33 4 1 Instrument Setup 33 4 2 Remote Control...

Страница 5: ...rement without Reflectors 54 7 4 Automatic Target Aiming ATR 55 7 5 SpeedSearch 56 7 6 Conformity to National Regulations 56 7 6 1 iCR70 56 7 6 2 CommunicationHandle 57 7 6 3 Dangerous Goods Regulatio...

Страница 6: ...to read and fully understand the fol lowing table with the different signal words and their definitions Supplemen tary safety information symbols may be placed within a warning message as well as supp...

Страница 7: ...stallations or simi lar situations Life Risk Precautions Local safety authorities and safety experts must be contacted by the per son responsible for the product before working in such conditions 1 4...

Страница 8: ...cinity of electrical installations such as power cables or electrical railways Precautions Keep at a safe distance from electrical installations If it is essential to work in this environment first co...

Страница 9: ...red Adhere to the regulations governing safety accident prevention and road traffic CAUTION Pointing product toward the sun Be careful when pointing the product toward the sun because the telescope fu...

Страница 10: ...your local passenger or freight transport company WARNING Exposure of batteries to high mechanical stress high ambient tempera tures or immersion into fluids This can cause leakage fire or explosion...

Страница 11: ...or WARNING Improperly repaired equipment Risk of injuries to users and equipment destruction due to lack of repair knowl edge Precautions Only Leica Geosystems authorised service centres are entitled...

Страница 12: ...se repetition frequency PRF 100 MHz Maximum average radiant power 0 33 mW Beam divergance 1 5 mrad x 3 mrad 16531_001 a a Laser beam 1 6 3 Distancer Measurements without Reflectors The EDM module buil...

Страница 13: ...rom a safety perspective class 3R laser products should be treated as poten tially hazardous Precautions Prevent direct eye exposure to the beam Do not direct the beam at other people CAUTION Reflecte...

Страница 14: ...d laser beam which emerges from the telescope objective The laser product described in this section is classified as laser class 3R in accordance with IEC 60825 1 2014 05 Safety of laser products Dire...

Страница 15: ...Reflected beams aimed at reflecting surfaces Potential hazards are not only related to direct beams but also to reflected beams aimed at reflecting surfaces such as prisms windows mirrors metallic sur...

Страница 16: ...red 20XX Made in Singapore iCR70 Type 16533_001 1 6 5 Automatic Target Aiming ATR The Automatic Target Aiming built into the product produces an invisible laser beam which emerges from the telescope o...

Страница 17: ...h 850 nm Maximum average radiant power 11 mW Pulse duration 20 ns 40 ns Pulse repetition frequency PRF 24 4 kHz Beam divergance 0 4 mrad x 700 mrad 16535_001 a a Laser beam 1 6 7 Electronic Guide Ligh...

Страница 18: ...dous for deliberate staring into the beam The beam may cause dazzle flash blindness and after images particularly under low ambient light conditions Description Value Wavelength 640 nm Maximum average...

Страница 19: ...force in this respect Leica Geosystems cannot completely exclude the possibility that other equipment may be disturbed CAUTION Use of the product with accessories from other manufacturers For example...

Страница 20: ...nal battery product to computer must be connected at both ends WARNING Use of product with radio or digital cellular phone devices Electromagnetic fields can cause disturbances in other equipment in i...

Страница 21: ...between the equipment and the receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technicia...

Страница 22: ...PVH0946 IC 5325A 0946 S No 1234567 This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this...

Страница 23: ...ce measurement Two measuring modes are available Prism mode This mode refers to the ability to meas ure distances to prisms Reflectorless mode This mode refers to the ability to measure distances with...

Страница 24: ...and iCR70 Target Snap A prism search method Snaps to the desired prism by ignoring other prisms which are known from the data base Angle measurement Distance measurement to prism Distance measurement...

Страница 25: ...enable software functions and applications which run on the device You can order new licences on myWorld or by contacting your local service partner 1 Store the licence key file key in the SYSTEM dir...

Страница 26: ...ted and removed Available capacity 1 GB and 8 GB USB stick All instruments have a USB port fitted as standard Internal memory All instruments have an internal memory fitted as standard Available capac...

Страница 27: ...f Instrument with tribrach and standard handle or CommunicationHan dle g GMP101 mini prism h GEB222 batteries i GFZ3 or GOK6 diagonal eyepiece j Counterweight for diagonal eyepiece k MS1 1 GB USB memo...

Страница 28: ...trument Components 16064_001 a b c e d k g f i j h a Carry handle b Optical sight c Telescope integrating EDM ATR EGL Speed Search d EGL flashing diode yellow and red e SpeedSearch transmitter f Speed...

Страница 29: ...16539_001 a b c d e a Compartment lid b USB stick cap storage c USB device port mini AB OTG d USB host port for USB stick e SD card port 16088_001 a b a CommunicationHandle b Communication side cover...

Страница 30: ...ment Press and hold for more than 5 s to reset the sensor Press and hold for more than 8 s to do a hard shutdown COM To toggle through the communication modes press the key repeatedly Communication mo...

Страница 31: ...et the sensor Press and hold for more than 8 s to do a hard shutdown COM To toggle through the communication modes press the key repeatedly Communication modes RS232 Bluetooth Long range Bluetooth LED...

Страница 32: ...r LED d Mode LED Description of the LED Indicators LED LED Status Description Power LED off Power is off green Power is on Link LED off No radio link to field controller red Radio link to field contro...

Страница 33: ...t Connect the instrument to the field controller To activate the laser plummet display the Compensator screen on the field controller Select Devices from the Home Menu Tap the arrow button to the righ...

Страница 34: ...ck push buttons and attach the handle Ensure that there is a tight fit with the instrument after releasing the push buttons If no connection can be found re check that handle is seated firmly 3 Swing...

Страница 35: ...amp hole 2 Attach the holder to the clamp using the clamp bolt 3 Adjust the angle and the height of the holder on the pole to a com fortable position 4 Tighten the clamp with the clamp bolt 5 Before p...

Страница 36: ...r the battery to become warm during charging Using the chargers recommended by Leica Geosystems it is not possible to charge the battery once the temperature is too high For new batteries or batteries...

Страница 37: ...d Click the battery into position 5 Place the battery housing into the battery compartment Push the battery housing in until it fits completely into the battery compart ment 6 Turn the knob to lock th...

Страница 38: ...urn the knob on the Communication side cover to the vertical posi tion to unlock the communication compartment 2 Open the lid of the communication compartment to access the com munication ports 3 To i...

Страница 39: ...the port 4 If desired store the lid of the USB stick on the underside of the com partment lid 5 Close the lid and turn the knob to the horizontal position to lock the compartment 4 7 Connecting to a...

Страница 40: ...d Press OK 1 Start the PC 2 Run the Setup_Leica_USB_XXbit exe to install the drivers necessary for Leica devices Depending on the version 32bit or 64bit of the operating system on your PC you have to...

Страница 41: ...ext 5 The Ready to Install the Program window appears 6 Install The drivers will be installed on your PC 7 The InstallShield Wizard Completed window appears 8 Click Finish to exit the wizard Operation...

Страница 42: ...his occurs because reflectorless measurements are made to the first surface returning sufficient energy to allow the measurement to take place For example if the intended target surface is the surface...

Страница 43: ...tiating a distance measurement the instrument sights the prism centre auto matically Vertical and horizontal angles and the distance are measured to the centre of the prism The lock mode enables the...

Страница 44: ...If the compensator and the horizontal corrections are activated in the instru ment configuration every angle measured in the daily work is corrected auto matically Check whether the tilt correction an...

Страница 45: ...trument should be protected from direct sunlight to avoid thermal warming It is also recommended to avoid strong heat shimmer and air turbulence The best conditions are early in the morning or with ov...

Страница 46: ...ion l t Compensator longitudinal and transversal index errors i Vertical index error related to the standing axis c Horizontal collimation error also called line of sight error ATR Hz ATR zero point e...

Страница 47: ...ace one The fine pointing has to be performed manually in both faces 6 Tap in the wizard to get to the next page Aim the telescope accurately at a target at about 100 m distant or less if not possible...

Страница 48: ...ach and onto a tripod 2 Turn on the instrument Connect the instrument to the field control ler Using the tribrach footscrews level the instrument with the elec tronic level Electronic level Select Dev...

Страница 49: ...circular level on the prism pole 4 a If the circular level is cen tred no adjustment is nec essary b If the circular level is not centred use an allen key to centre it with the adjustment screws Afte...

Страница 50: ...descri bed by the centre of the laser dot should not exceed 3mm at a height of 1 5m If the centre of the laser dot describes a perceptible circular move ment or moves more than 3 mm away from the poi...

Страница 51: ...n the User Manual particularly after the product has been dropped stored for long periods or transported 6 2 Storage Respect the temperature limits when storing the equipment particularly in summer if...

Страница 52: ...a temperature not greater than 40 C 104 F and clean them Remove the battery cover and dry the battery compartment Do not repack until everything is completely dry Always close the transport container...

Страница 53: ...2000 7000 360 Mini prism GRZ101 450 1500 800 2600 1000 3300 Mini prism GMP101 800 2600 1200 4000 2000 7000 Reflector tape GZM31 60 mm x 60 mm 150 500 250 800 250 800 Shortest measuring distance 1 5 m...

Страница 54: ...urement without Reflectors R500 Kodak Gray Card Range D Range E Range F m ft m ft m ft White side 90 reflective 300 980 500 1640 500 1640 Grey side 18 reflective 150 490 250 820 250 820 Range of Measu...

Страница 55: ...ng distance 360 prism Target aiming 1 5 m Shortest measuring distance 360 prism Target locking 5 m Type Accuracy ATR angle accuracy Hz V std dev ISO 17123 3 1 0 3 mgon Direction of prism movement Tang...

Страница 56: ...on Definable search win dows Yes Type Description Principle Digital signal processing Type Infrared laser 7 6 Conformity to National Regulations 7 6 1 iCR70 FCC Part 15 applicable in US Hereby Leica G...

Страница 57: ...radio equipment type CCD6 RH16 RH17 is in compliance with Directive 2014 53 EU and other applicable European Directives The full text of the EU declaration of conformity is available at the fol lowing...

Страница 58: ...e Leica products can be transported correctly Damaged or defective batteries are prohibited from being carried or transported onboard any aircraft Therefore ensure that the condition of any battery is...

Страница 59: ...cant influence on the automated target In addition the selec ted EDM mode affects the ATR performance Under good environmental condi tions and with a clean properly aligned target the accuracy of the...

Страница 60: ...munication data transfer This port is located at the base of the instrument CommunicationHandle Hotshoe connection for Communication Handle This port is located on top of Communi cation side cover Blu...

Страница 61: ...2 not used c Pin 3 Single ground d Pin 4 RxD RS232 receive data In e Pin 5 TxD RS232 transmit data Out 16547_001 88 mm 76 mm 164 mm 177 mm 226 mm 195 7 mm 336 mm Pin Assignments of the 5 Pin LEMO 0 P...

Страница 62: ...is of instrument Accuracy Deviation from plumb line 1 5mm 2 sigma at 1 5m instrument height Diameter of laser point 2 5mm at 1 5m instrument height Endless horizontal and vertical drives Type Descript...

Страница 63: ...g The effects of condensation are to be effectively counteracted by periodically drying out the instru ment Type Additive Constant mm ATR SpeedSearch Standard prism GPR1 0 0 yes yes Mini prism GMP101...

Страница 64: ...air pressure Air temperature Relative humidity For highest precision distance measurements the atmospheric correction should be determined with an accuracy of 1 ppm The following parameters must be r...

Страница 65: ...for DD2 are always negative and are derived from the following for mula TS_106 D2 H R 106 DD2 Reduction to mean sea level ppm H Height of EDM above sea level m R 6 378 106 m The magnitude of the proje...

Страница 66: ...e F air pressure inch Hg and height ft at 60 relative humidity TS_109 130 F 120 F 110 F 100 F 90 F 80 F 70 F 60 F 50 F 40 F 30 F 20 F 10 F 0 F 10 F 20 F 130 F 120 F 110 F 100 F 90 F 80 F 70 F 60 F 50...

Страница 67: ..._113 X B Y2 Horizontal distance m Height difference m Y sinz X cosz z Vertical circle reading A 1 k 2 R 1 47 10 7 m 1 B 1 k 2 R 6 83 10 8 m 1 k 0 13 mean refraction coefficient R 6 378 106 m radius of...

Страница 68: ...km AC Additive constant of the reflector m TS_112 Y A X Y TS_113 X B Y2 Horizontal distance m Height difference m Y sinz X cosz z Vertical circle reading A 1 k 2 R 1 47 10 7 m 1 B 1 k 2 R 6 83 10 8 m...

Страница 69: ...e read and accepted the terms and conditions of the Leica Geosystems Software Licence Agree ment Installation or use of the software or any part thereof is deemed to be an acceptance of all the terms...

Страница 70: ...ystems AG Heinrich Wild Strasse CH 9435 Heerbrugg Switzerland Phone 41 71 727 31 31 www leica geosystems com 871929 1 0 0en Original text Published in Switzerland 2018 Leica Geosystems AG Heerbrugg Sw...

Отзывы: