4.
•
Aim the telescope accurately at a target at about 100 m distant
or less if not possible. The target must be positioned at least
27°/30 gon above or beneath the horizontal plane.
•
Start the procedure in face I.
•
Press the measurement keys to measure and to continue to the
next step.
Tap
in the wizard to get to the next page.
☞
The fine aiming has to be performed manually in both faces.
5.
Adjustment Accuracy
After pressing the last in the wizard the results are shown and
stored to the instrument.
6.
Tap to get back to the
Devices
page.
The purpose of the calibration report is a documentation of the results of the
field calibration. The report proofs the quality of the equipment for quality
insurance.
At the end of the sensor calibration, a report (*.calibration) is created auto-
matically. The report contains all calibration values.
The report is stored on the hard drive of the controller (for instruments with
keyboard unit only):
The reports and results of past calibrations can also be exported.
Calibration report
Check & Adjust
39