20 Close the lamp door securely.

21 Hold the aspherical lens in the light port and turn the base right side up again.

22 Drop the mounting ring into the light port so it surrounds the aspherical lens.

23 Using the ball end of the 2.5mm wrench, press one of the opened ends of the mounting ring

into the groove around the circumference of the lens. Put tissue on one of your fingers and

hold this end of the mounting ring into the groove. 

24 Using the other hand with the ball end of the 2.5mm wrench, run the ball around the 

circumference of the lens pressing the mounting ring into the groove along the way.

Do not touch the lens surface with your fingers.

25 Once the mounting ring is in the groove, use an air syringe or slightly blow on the aspherical

lens to remove any dust. Using a lens tissue if necessary, remove any fingerprints or marks.

26 Lift the stand and reposition it over the base, making sure to align the back plate with the 

stand’s groove.

27 Tip the microscope on its back holding the base with one hand.

28 Remount the four rubber feet with screws and tip the microscope back onto the base, 

returning it to it’s normal operating position.

29 Reconnect the power cord into the back of the microscope and rettach the cord wrap / 

lock if necessary.

30 Position the field diaphragm with protective window onto the light port so the 

holes in the stand align with the holes in the field diaphragm and the lever for the 

field diaphragm assembly is facing you.

31 Using the 2.5mm wrench and the mounting screws provided, attach the field diaphragm 

to the microscope.

32 Lower the condenser bracket with the condenser focus.

33 Install the condenser into the condenser bracket so the numbers on the condenser are 

facing forward and the pin on the back of the condenser is aligned with the slot in the 

condenser bracket.

34 Screw the condenser centering screws into the condenser bracket simultaneously 

until the pin in the rear of the condenser is half way into the notch.

35 Raise the condenser to the top position with the condenser focus

4

Содержание DM E

Страница 1: ...Leica DM E Koehler Illumination Instruction Manual ...

Страница 2: ...r Phase Turret condensers 2 Lens in holder 3 Mounting bolts 4 Koehler field diaphragm 5 Mounting ring 6 Aspherical lens 7 DM E light port 8 Condenser centering screws 2 9 2 5mm Allen wrench 10 Blue daylight and green filters included with Koehler Illumination Kit ...

Страница 3: ...inate stray light from entering the field of view with the use of a field diaphragm This is done by a combination of precision optics condenser adjustment and adjusting a field diaphragm so only the field of view is illuminated The proper adjustment for Koehler illumination is described after the installation instructions Check that you have the following components in your Koehler illumination Ki...

Страница 4: ... port window Save this for possible later use 11 Remove the power cord from the back of the instrument 12 Carefully tip the microscope on its back and remove the 4 screws holding the rubber feet 13 While holding the base in place with one hand tip the microscope back onto its base to its normal operating position 14 Carefully lift the microscope off of its base 15 Looking into the light port you w...

Страница 5: ...k plate with the stand s groove 27 Tip the microscope on its back holding the base with one hand 28 Remount the four rubber feet with screws and tip the microscope back onto the base returning it to it s normal operating position 29 Reconnect the power cord into the back of the microscope and rettach the cord wrap lock if necessary 30 Position the field diaphragm with protective window onto the li...

Страница 6: ...wing through the microscope close the field diaphragm which you have just installed using the field diaphragm lever You will see the iris leaves imaged within the field of view 10 Using the condenser adjustment focusing knob focus the condenser until the leaves of the field diaphragm are in sharp focus 11 Center the image of the field diaphragm by turning the condenser centering screws This is bes...

Страница 7: ...6 ...

Страница 8: ...olicy of continuous development we reserve the right to change specifications without notice Aufgrund der fortlaufenden Neuentwicklungen behalten wir uns das Recht vor jederzeit ohne Vorankündigung Änderungen vornehmen zu können En raison de notre politique de développement continu nous nous réservons le droit de modifier les spécifications sans préavis En vista de nuestra política de desarrollo c...

Отзывы: