ScanStation C10/C5, Setting Up the Instrument
47
3. Place the instrument on the tribrach (e) and secure it with the tribrach’s locking
knob.
4. Turn on the instrument by pressing the ON/OFF button (f). Go to
Status
,
Level
and Laser Plummet
,
Plummet
and activate the laser plummet (g).
5. Move the tripod legs (a) and use the tribrach footscrews (h) to centre the
plummet (i) over the ground point.
6. Adjust the tripod legs (j) to level the circular level (k).
7. By using the electronic level (
Status
,
Level and Laser plummet
,
Level
) turn the
tribrach footscrews (h) to precisely level the instrument.
8. Centre the instrument precisely over the ground point (i) by shifting the tribrach
on the tripod plate.
9. Repeat steps 7. and 8. until the required accuracy is achieved.
Please see also section "Scanning with ScanStation C10/C5" in the
Cyclone
documentation for more information.
Содержание ScanStation C10
Страница 1: ...Leica ScanStation C10 C5 User Manual Version 5 0 English ...
Страница 41: ...ScanStation C10 C5 Description of the System 41 ...
Страница 85: ...ScanStation C10 C5 Troubleshooting 85 ...
Страница 97: ...ScanStation C10 C5 Care and Transport 97 ...
Страница 118: ...118 ScanStation C10 C5 Safety Directions a Laser beam b Exit for laser beam b C10_032 a ...
Страница 126: ...126 ScanStation C10 C5 Safety Directions Labelling GEB271 ...
Страница 127: ...ScanStation C10 C5 Safety Directions 127 Labelling GKL271 ...
Страница 149: ...ScanStation C10 C5 Microsoft End User License Agreement EULA 149 ...
Страница 152: ......