2.3. 

Control system  

The  colour  LCD  touch  screen  display  allows  the  user  to  easily  see  operating  conditions  and  to  alter 
settings through icon based menus. 

Software  controlled  PID  loops  constantly  monitor  temperature  and  CO2  values,  making  automatic 
adjustments to keep the chamber atmosphere within limits. 

The  controller  is  equipped  with  a  built  in  feature  that  logs  incubator  data  directly  to  a  removable  SD 
card. This data is in Excel format, making it easy to view with a PC. 

 Section 7

 

A RS232 serial port is fitted as standard. This allows you to connect the incubator to a PC so that you 
can remotely access the SD card data. PC software is required to do this. Contact your dealer or LEEC 
directly for advice. 

Other optional communication options are Ethernet PC interface and 4-20mA retransmit. 

 Section 10

 

2.4. 

CO

Sensor 

CO

is  measured  and  controlled  using  a  highly  accurate  infra-red  sensor.  This  sensor  remains 

practically drift free between calibration periods.  

The sensor is not mounted within the chamber; instead a small pump passes the chamber atmosphere 
through the sensor.  

The sensor does not have to be removed during high temperature sterilisation. 

 

2.5. 

Diagrams 
 

2.5.1. 

Outer view 

 

 
Outer door 

 

 

Door latch

 

 

 

 

Magnetic locks 

 

 

 

 

 

Adjustable feet 

    Inner glass door 

 

 

Содержание Culture Safe Touch 190

Страница 1: ...190 Touch 190S Address LEEC Ltd Private Road No 7 Colwick Industrial Estate Nottingham NG4 2AJ UK Tel 44 115 961 6222 Fax 44 115 961 6680 Enquiries E mail general leec co uk Sales E mail sales leec co...

Страница 2: ...G 3 3 STORAGE 4 INSTALLATION AND SET UP 4 1 LOCATION AND AMBIENT CONDITIONS 4 2 SHELVES AND USABLE SPACE 4 3 WATER TRAY 4 4 GAS CONNECTION 4 5 ELECTRICAL CONNECTION 5 GETTING STARTED 5 1 START UP 5 2...

Страница 3: ...CLE 9 ERROR MESSAGE SCREENS 10 OPTIONS 10 1 RS 232 SERIAL INTERFACE 10 2 BMS VOLT FREE ALARM RELAY 10 3 ETHERNET INTERFACE 10 4 4 20mA RETRANSMIT 10 5 MULTI DOORS 11 ACCESSORIES 12 CLEANING 12 1 OUTER...

Страница 4: ...the incubator are subject to change without notice 1 2 Intended use Culture Safe Touch CO2 incubators are intended for mammalian cell cultivation under typical conditions of 37 0 C The pH of the cell...

Страница 5: ...Only original spare parts from LEEC can be used There are no user serviceable parts within the incubator LEEC takes no responsibility for 3rd party accessories or equipment used with this incubator DA...

Страница 6: ...king and water tray are also made from stainless steel The outer housing is made from stainless steel and powder coated with anti bacterial biocide paint 2 2 Doors The outer door has a heater on its i...

Страница 7: ...ted as standard This allows you to connect the incubator to a PC so that you can remotely access the SD card data PC software is required to do this Contact your dealer or LEEC directly for advice Oth...

Страница 8: ...CO2 sensor return Chamber PT100 Shelves Left hand racking Right hand racking Service port CO2 sensor intake Inner door gasket Water tray 2 5 3 Rear connections Ethernet CO2 Inlet BMS relay Sampling po...

Страница 9: ...wing items Contact your dealer with details of any shortages Standard T190 accessory pack Qty Stainless steel perforated shelves 4 Shelf racking left and right hand 1 set Racking support bar 1 Stainle...

Страница 10: ...he set point and ambient temperatures is less than 6 0 C then the set point can be exceeded and temperature fluctuation will occur Our performance data was produced at the ideal ambient condition of 2...

Страница 11: ...m base 70mm For practical purposes this is the perforated area on each shelf Do not place any samples outside of this space Do not over fill each shelf Allow sufficient room between samples to allow f...

Страница 12: ...s with the correct fittings and other gas accessories can be supplied by LEEC Section 10 For correct connection of CO2 follow these steps in order 1 Connect the supplied tubing and HEPA filter to the...

Страница 13: ...l supply of 230V AC 10 50 60Hz This incubator MUST be connected to a protective earthed supply Further details can be found in the technical data Section 15 1 Plug the supplied power lead into the IEC...

Страница 14: ...ay This screen will be shown Use the touch screen to enter the correct time Press to confirm selection Repeat these steps for setting the date 5 3 Factory presets The following values are preset to al...

Страница 15: ...re set point CO2 process value Chamber CO2 set point Sterilise cycle start Menu escape 5 5 Changing temperature set points and high low alarms To change the temperature set point touch this area of th...

Страница 16: ...high alarm will trigger at 37 5 C Use the keypad to change the high alarm threshold Press to confirm selection Use the keypad to change the low alarm threshold This must be negative so remember to fi...

Страница 17: ...um set point is limited to 0 0 Press to confirm selection If the CO2 set point is 0 0 then CO2 control is disabled Use the keypad to change the high alarm threshold Press to confirm selection Use the...

Страница 18: ...nd the alarm status screen will show the corresponding alarm will in red For instance this shows high temperature alarm condition Pressing this will mute the alarm The alarm is now acknowledged and th...

Страница 19: ...st Graphs are only stored while the unit is powered This feature is independent of the SD data logging function Section 7 On the main screen press the graph symbol The graph menu opens Choose which gr...

Страница 20: ...en the door is opened The door icon turns red as soon as the door is opened After a set time the alarm will sound This can be muted in the alarm status menu Section 5 7 This delay can be changed by pr...

Страница 21: ...ould you require any assistance from LEEC 6 5 Historical event log The controller has a built in feature that displays the time and date of 100 historical events When 100 events have been logged the o...

Страница 22: ...a user name and password To enter a user name and password press the User icon Press here on Line 1 to enter user name Press to save Then press here to enter password Press to save The user name and p...

Страница 23: ...ard is pre formatted to work with the controller s operating system Only cards supplied by LEEC can be used Using any other type of card will not work with the incubator and will cause the controller...

Страница 24: ...lity Menu The Status screen will state that the card is removed Insert the SD card into the slot to the right of the controller label side closest to controller The SD Card status screen shows that th...

Страница 25: ...n easily change the filename to start a new log Press the filename on the screen in this case 12345 Enter a new file name in this case 67890 Press to store selection Press to confirm the new filename...

Страница 26: ...removing the SD Card Failure to follow these steps will results in corrupted data On the SD Card status screen press the SD card tools icon Press the remove card icon Press the confirm icon It is now...

Страница 27: ...aises the chamber temperature to 200 C for 2 hours Independent evaluation has shown this to be effective against all spores and bacteria commonly associated with laboratory incubator contamination 8 1...

Страница 28: ...are now complete Press to proceed This shows that the door locks have activated and that Operator 6 is logged on Press to proceed 8 3 Sterilise cycle The sterilise cycle has now started A temperature...

Страница 29: ...the chamber temperature reaches 38 C the door locks are deactivated and CO2 control is enabled The time taken for a full sterilisation cycle is dependent on the cool down phase As cooling relies on t...

Страница 30: ...are locked during this phase CO2 control is disabled The chamber temperature is displayed The cooling phase ends once the temperature falls below 38 0 C Press to continue to main screen 8 5 Aborting...

Страница 31: ...d has failed Contact LEEC Service immediately Door probe failure Door probe is out of limits and has failed Contact LEEC Service immediately SD card critical error SD has not been safely removed Press...

Страница 32: ...rm condition clears The connections are as shown below 10 2 4 20mA retransmission RETX This is an option that must be specified at the point of ordering This feature allows the chamber temperature and...

Страница 33: ...uctions will be sent separately when this option is ordered 10 5 Multi door 8ID This can be retro fitted The inner glass door can be replaced with a multi door This has 8 smaller glass doors in a stai...

Страница 34: ...a sterile cloth so as not to introduce any contamination Never use any of the following Any chlorine based detergents or bleaches Any acidic cleaning agents Sodium Azide Iodine Ferric Chloride Wipe ov...

Страница 35: ...l distributor in the first instance Returns to LEEC will only be accepted with prior authorisation from the Service Department We recommend that you keep all packaging When returning a unit please inc...

Страница 36: ...ELECTRICAL DATA Nominal voltage 50 60Hz V 230 1N PE Power T190 W 380 Power T190S sterilise cycle W 1800 Power consumption kWh 0 058 Main fuse F1 T190 6 3A 6 3 x 32mm 250V TT Main fuse F1 T190S 10A 6...

Страница 37: ...Jan 2012 T190 Iss 2 Page 37 of 38 15 2 Dimensional drawings...

Страница 38: ...Jan 2012 T190 Iss 2 Page 38 of 38 16 EC Declaration of Conformity...

Отзывы: