6
PP010-OM-E Rev A
Specifications
Electrical
Characteristics
Bandwidth
DC to 200 MHz (
±
3 dB)
Attenuation
10:1
±
3%
Input RC
10 M
Ω
±
2% // 12.5 pF
±
2 pF
Compensation Range
18-35 pF
Probe Length
1.5 m
Connector
Type
BNC
Environmental
Conditions
Temperature (operating)
5 to 40 °C
(non-operating)
-20 to +70 °C
Humidity (operating)
80% RH (5 to 31 °C)
Altitude (operating)
up to 2000 m (6562 ft)
Electrical
Ratings
Maximum Input Voltage
600 V (DC + AC
peak
)
Note:
The maximum input voltage derates when frequency response is 1 MHz
or higher. See voltage derating curve on the next page.
EC Declaration of Conformity
The PP010 probe meets intent of the European Council Directive 73/23/EEC
for product safety per the following standard:
EN 61010-031:2002
Safety requirements for hand-held
probe assemblies for electrical test
and measurement
Installation
(Overvoltage)
Category I (All mains isolated
terminals)