background image

 

PP010-OM-E Rev A

 3 

Introduction 

The PP010 is a passive probe to be used with a LeCroy WaveJet oscilloscope 
having an input impedance of 1 M

. It has an attenuation ratio of 10:1. 

Safety Information 

This symbol appears on the product: 

 

It refers you to additional information contained in this manual. The 
corresponding information in the manual is similarly denoted. 
To avoid personal injury and to prevent fire or damage to the probe or any 
products connected to it, review and comply with the following safety 
precautions. 

 WARNING -- ELECTRIC SHOCK HAZARD 
Do not remove the probe case.

 Removal of the probe case may expose 

you to electric shock. 

Do not disassemble the probe. 

Disassembling the probe may expose 

you to electric shock. 

Do not disconnect the probe from the scope during measurement.

 

Disconnecting the probe during measurement may expose you to 
electric. 
Connect the probes grounding terminal to the ground potential of the 
device under test. Failure to connect to ground potential may expose you 
to electric shock. 

Connect to properly grounded instruments

. This probe is to be used only 

with test instruments having input connectors with the BNC shield (ring) 
connected to earth ground. 

Connect the probe properly: 

Connect the probe to the measurement 

instrument before connecting the probe to test circuit. Do not disconnect the 
probe from the test instrument while the input is connected to the test circuit. 
Do not connect the probe ground terminal to any point in the test circuit that is 
at a potential other than earth ground. 

 

Do not apply to the input any potential that exceeds the maximum 

ratings of the probe.

 Refer to specifications. 

Do not use in wet or explosive atmospheres. 
For indoor use only.

 This probe is intended for indoor use and should be 

operated in a clean, dry environment.

 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание PP010

Страница 1: ...LeCroy PP010 Passive Probe Instruction Manual Revision A February 2006 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ... warranty with transportation charges prepaid This warranty replaces all other warranties expressed or implied including but not limited to any implied warranty of merchantability fitness or adequacy for any particular purposes or use LeCroy shall not be liable for any special incidental or consequential damages whether in contract or otherwise 2006 by LeCroy Corp All rights reserved LeCroy Active...

Страница 3: ...pose you to electric Connect the probes grounding terminal to the ground potential of the device under test Failure to connect to ground potential may expose you to electric shock Connect to properly grounded instruments This probe is to be used only with test instruments having input connectors with the BNC shield ring connected to earth ground Connect the probe properly Connect the probe to the ...

Страница 4: ...y The probe tip is extremely sharp Use care when handling to prevent injury including accidental skin puncture Use of the probe and or the test instrument it is connected to in a manner not specified by the manufacturer may impair the protection mechanisms Components www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 5: ...nt Kit PKIT2 5MM 101 contains the following items Part Quantity Standard Ground Lead 1 Sprung Hook 1 Color Coding Rings set 1 Adjustment Tool 1 Ground Attachment 1 IC Test Tip 1 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 6: ...aximum Input Voltage 600 V DC ACpeak Note The maximum input voltage derates when frequency response is 1 MHz or higher See voltage derating curve on the next page EC Declaration of Conformity The PP010 probe meets intent of the European Council Directive 73 23 EEC for product safety per the following standard EN 61010 031 2002 Safety requirements for hand held probe assemblies for electrical test ...

Страница 7: ...00 1000 0 1 1 10 100 1000 Frequency MHz Voltage V DC AC peak Typical Input Impedance 10 100 1000 10000 100000 1000000 10000000 0 01 0 1 1 10 100 1000 Frequency MHz Zin ohms www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 8: ...en first connecting a passive probe to the oscilloscope input LF compensation is performed by connecting the input of the probe to a low frequency square wave such as the oscilloscope calibrator signal set to 1 kHz The compensation is adjusted by rotating the trimmer pot accessible through the small hole in the center of the housing near the BNC connector Use the tool supplied with the probe for t...

Отзывы: