5. PICTURE DISPLAY
89
SIZE = REAL
Figure 5-33 Selecting the display size
5.8.2
Gamut Error Display
To display the locations of gamut errors and luminance errors on the picture, follow the
procedure below.
These errors are defined by the ranges that you specify by setting Gamut Upper and Gamut
Lower, Composite Upper and Composite Lower, and Luminance Upper and Luminance
Lower in the status menu. If Gamut Error, Composite Gamut Error, or Level Error is set to
OFF, the corresponding errors are not displayed.
Reference Gamut Upper/Lower, Composite Upper/Lower
→ Section
8.2.3, “Error Setup 3”, Section 8.2.4,
“Error Setup 4”
Procedure
PIC
→
F•7
DISPLAY →
F•2 GAMUT ERR DISP: OFF / WHITE / RED / MESH
Settings
OFF:
Gamut errors are not displayed.
WHITE:
The picture intensity is halved, and gamut error are marked in white.
RED:
The picture intensity is halved, and gamut error are marked in red.
MESH:
Gamut errors are marked with a mesh pattern.
5.8.3
Turning the Information On and Off
To turn on and off the display of the following information that you arranged in the layout,
follow the procedure below.
This setting is valid on the display that appears when PIC is pressed. For multi display and
other displays, it is fixed to ON.
• Sub tab items (FORMAT, INPUT, TIME, DATE)
• Option tab options (Format, Input, Time)
Procedure
PIC
→
F•7
DISPLAY →
F•4 STATUS INFO: ON / OFF
Содержание LV 5480
Страница 15: ...2 VIDEO SIGNAL WAVEFORM DISPLAY 7 SCALE UNIT HDV SDV SCALE UNIT HD SD SCALE UNIT 150 ...
Страница 38: ...2 VIDEO SIGNAL WAVEFORM DISPLAY 30 COLOR MATRIX XYZ COLOR MATRIX GBR COLOR MATRIX RGB ...
Страница 98: ...5 PICTURE DISPLAY 90 STATUS INFO ON Figure 5 34 Turning the information on and off ...
Страница 119: ...7 AUDIO DISPLAY 111 DISPLAY MODE METER DISPLAY MODE SURROUND DISPLAY MODE STATUS Figure 7 8 Selecting the display mode ...
Страница 163: ...8 STATUS DISPLAY 155 DISPLAY SERIAL DISPLAY COMPO DISPLAY BINARY Figure 8 10 Selecting the display format ...
Страница 202: ...9 EYE PATTERN DISPLAY SER02 SER09 194 SWEEP 2UI SWEEP 16UI Figure 9 9 Selecting the sweep time ...