XI. Fault Analysis
Phenomenon
Cause analysis
Treatment
Power director does not
show
Without power
Inspect the plug
Fuse broken
Change fuse
Temp. Controller
shows“
□□□□
”
Sensor broken
Change sensor
Controller broken
Change controller
Evaporator frost or
Chamber with frost
Open the door
frequently when doing
the Low-Temp. test
Over 50 degree make dry and
reduce open door
Left hole with bad seal
Inject the inner hole by rubber
Door opening
Close the door
Hard to make the Temp.
down
Evaporator frost
Dry the chamber
Environment Temp. too
high
Down environment Temp.
Fan works or not
Check fuse and fan
Compressor works or not
Change Compressor
Compressor works but not
refrigeration
Check the cryogen
Check the ice or oil block
Specification mixed
Setting correct and restart
Temp. up continue
Evaporator frost
Dry the chamber
Abnormal Knocking
Circle Fans loose
Check and adjust
Condenser, Fan,
compressor loose
Check and adjust or contact with
us
Bad evenness degree
Sample hot
Reduce sample quantity
Evaporator frost
、
wind
block
Dry the chamber
、
Restart
Controller instability
Power do not match
Change the Power
Voltage instability
Make sure stability voltage
Hard make Temp. up
Over-Temp. setting too
low
Adjust the Temp. correct
Meter setting too low
Set the Temp. correct
Meter heating director on
but without input function
Change the meter
Meter heating but the
heater doesn’t work
Change the heater
Fan doesn’t work
Change fan or fuse
Sensor broken
Change sensor
Temp. over bigger than
setting
Meter setting incorrect
Set again and the manual
Heater working do not
stop
Change Controller