Working conditions 

 
1. Environment Temp.: 5

32

 (if setting temp.

10

environment

28

). 

2. Humidity

80%RH    3. Air Pressure

 80-106Kpa 

4. There must be no shaking of the incubator and no caustic air around the chamber. 
5. Should avoid light and cooling environment. 
6. Keep distance of dust, install horizontally and make sure there is distance between the equipment and wall. 
 
 

Safety information 

 
1. After you receive the equipment, to avoid the damage of the compressor during a long transportation period, 
please keep the machine without moving during one day before using it. To make sure equipment safety, please 
install outer protection and also supply power as the equipment requires. 
2. Do not test the flammable, combustible, poisonous or corrosive material. 
3. The equipment should be installed horizontally. 
4. Only trained personnel can disassembly and repair the equipment. 
5. If you add one working equipment into the chamber, if the power is bigger than 2A, plug the power outside, 
do not use this equipment power. 
6. When the chamber temp.

50 

, please do not set the low temp. settings to make sure the compressor can 

be used for a long time. 
7. Anyone who operates the equipment must first read this manual 
 

Operation cautions  

 
1. There is a test hole at the top of the incubator. If you add another equipment inside the chamber, please do it 
through this hole. 
2.

 

If there is no need for U.V. light, turn off the U.V. lamp to avoid influence on temperature. 

3.

 

Before operating the cooling function, please dry the chamber for 1 hour under 50 ºC. 

4.

 

The chamber has a vertical air system. Please do not fully fill the shelfs with materials, the test material must 
occupy less than 1/3 of the shelf. 

5.

 

Once the test environment > 35

 or Use Temp. > 50

, do not set the low temp. setting 

6.

 

Please clean the chamber and surface of the machine with pure water. When the machine is not used, please 
plug off the electricity, and make sure the inside and outside is clean and dry. 

 

 

Содержание INCR-070-001

Страница 1: ...LBX INC R Refrigerated Incubator Please read the User Manual carefully before use and follow all operating and safety instructions user manual english ...

Страница 2: ...ith the following information Serial number Description of the problem Your contact information Warranty This instrument is guaranteed to be free from defects in materials and workmanship under normal use and service for a period of 24 months from the date of invoice The warranty is extended only to the original purchaser It shall not apply to any product or parts which have been damaged on accoun...

Страница 3: ...licon 7 50mm diameter test hole Product specifications Model INCR 070 001 INCR 150 001 Voltage 220 240V 50Hz 60Hz Temp Range 0 65 Temp Motion 0 5 Temp Uniformity 1 Cryogen R134a Power 800W 1000W Inner Chamber size cm 35 x 41 5 x 50 40 x 49 x 75 Exterior Size cm 56 x 57 x 103 62 x 64 x 129 Inner Chamber Temp Distribution map 9 7 8 4 1 3 2 6 5 Temperature Graph 30 20 10 0 30 60 10 20 30 40 50 60 30 ...

Страница 4: ...you add one working equipment into the chamber if the power is bigger than 2A plug the power outside do not use this equipment power 6 When the chamber temp 50 please do not set the low temp settings to make sure the compressor can be used for a long time 7 Anyone who operates the equipment must first read this manual Operation cautions 1 There is a test hole at the top of the incubator If you add...

Страница 5: ... 30 seconds press the down keyboard for 4 seconds the system will start working again when the buzzer sounds please click any keyboard to stop the sound 3 Illumination click the illumination button and then it will turn on click again and it will close Without timer setting With timer setting ⅳ Inner Chamber Specification setting Press the set button for more than 3 seconds the screen will show LC...

Страница 6: ...uP and compressor delay time out then compressor begin to work 50 0 50 0 0 2 Dn Cooling refrigeration backlash compressor in manual model when PV SP dn the compressor shut down 51 0 uP 0 1 0 T Control cycle 1 60sec 5 P Proportional band 1 0 measure value 35 I Integration time 1 1000sec 200 d Differential time 1 1000sec 200 Pb 0 adjust Sensor zero display error adjusts Pb actual temp display number...

Страница 7: ...n is pressed it will exit settings automatically and will not keep the specification Paramet er Name Function Range factory value Lc Password when Lc 18 we can check the ambient temperature Hd Ambient temp Parameter Parameter Function range Factory value Lc Lc 123 can see and modify each parameter S H When ambient temp SH Temp Setting value the compressor normally open operation On the contrary th...

Страница 8: ...en Check the ice or oil block Specification mixed Setting correct and restart Temp up continue Evaporator frost Dry the chamber Abnormal Knocking Circle Fans loose Check and adjust Condenser Fan compressor loose Check and adjust or contact with us Bad evenness degree Sample hot Reduce sample quantity Evaporator frost wind block Dry the chamber Restart Controller instability Power do not match Chan...

Страница 9: ... vertederos De conformidad con la Directiva 2002 96 CE los usuarios de la Unión Europea de aparatos eléctricos y electrónicos tienen la oportunidad de retornar el instrumento para su eliminación al distribuidor o fabricante del equipo después de la compra de uno nuevo La eliminación ilegal de aparatos eléctricos y electrónicos es castigada con multa administrativa Les appareils électriques et élec...

Страница 10: ...www labbox com ...

Отзывы: