Thermocouple Input Specifications
Calibration
NIST Monograph 125 (IPTS-68)
Overall error at 25°C
±0.01% of full span ± conformity error
Span tempco
±0.003% of reading/°C
Reference junct. accuracy
0.5°C, 10°C to 40°C
Input resistance
1 Gohm
Input current
100 pA
Max lead resistance
1 kohm max for rated accuracy
Overvoltage protection
125 Vac
NMR at 50/60 Hz
80 dB plus selectable digital filter
CMR, DC-60 Hz
120 dB with 500 ohm imbalance
CMV, DC-60 Hz
250 Vac from power and earth grounds