- 32 -
ALARMS
: the load cell is not connected or is incorrectly connected; the load cell signal exceeds
39 mV; the conversion electronics (AD converter) is malfunctioning; the load cell is a 4-
wire and there are no jumpers between EX- and REF- and between EX+ and REF+.
: the weight display exceeds 110% of the full scale.
: internal instrument converter failure; check load cell connections, if necessary contact
technical assistance.
: the weight exceeds the maximum capacity by 9 divisions.
: maximum displayable value exceeded (value higher than 999999 or lower than -999999).
: weight too high: zero setting not possible.
: this message appears in the sample weight setting, in real calibration, after the fifth
sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press ESC to quit the
setting mode leaving the previous value unchanged. Examples: a number of decimals is
selected for full scale which exceeds the instrument's display potential; value above the
maximum setting value; the weight value set in sample weight verification does not match
the detected mV increase.
:
lock active on menu item, keypad or display.
: It’s not possible to display properly the number because is greater than 999999 or less
than -999999.
: buffer battery low, loss of date and time of Real-Time Clock. Confirm with ENTER to
continue; leave the instrument on for at least 12 hours to charge the battery, if the alarm
persists contact technical assistance.
: an incorrect date has been detected: go into the related menu to check and correct it.
: gross weight equal to zero: the semi-automatic tare operation cannot be performed.
Serial protocol alarms:
MODE
Bit LSB
76543210
xxxxxxx1
76543210
xxxx1xxx
76543210
xxxxxx1x
76543210
xxxxx1xx
76543210
On gross:
xxx1xxxx
On net:
xx1xxxxx
The response to the
zero command is a
'value not valid' error
(error code 3)
Status
Register
MODBUS RTU
ASCII
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_ &aa#CR
RIP *
__O-F_
__O-L_
__O-F_
__O-L_
__O-F_
__O-F_
HDRIP-N
_ERCEL
_ER_OL
_ER_AD
######
_ER_OF O__SET
CONTIN
_ERCEL
_ER_OL
_ER_AD
^^^^^^
_ER_OF O__SET
* For RIP remote displays, if the message exceeds 5 digits the display reads
.