background image

- 11 - 

MAXIMUM CAPACITY 

 



: Maximum displayable weight (from 0 to max full scale; default: 0). When the weight exceeds 

this value by 9 divisions, the display shows 



. To disable this function, set 0. 

 
 

TARE WEIGHT ZERO SETTING 

 

This menu may also be accessed directly from the weight display, holding down the 

0

 

key for 3 seconds. 
 
Perform this procedure after having set the THEORETICAL CALIBRATION data. 
 

Use this function to set to zero the weight of the empty system after commissioning and then later 
on to compensate zero variations due to the presence of product residues. 
Procedure: 
 

-

 

Confirm the message 



 (Zero) by pressing ENTER. 

-

 

The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing. 

-

 

Confirming once again, the weight is set to zero (the value is stored to the permanent memory). 

-

 

Press 

 to display the value of the total weight reset by the instrument, given by the sum of all 

of the previous zero settings. 

 
 

ZERO VALUE MANUAL ENTRY 

 

WARNING:

 Perform this procedure only if it’s not possible to reset the weighed structure tare, for 

example because it contains product that can not be unloaded. 

 
Set in this parameter the estimated zero value (from 0 to max 999999; default: 0). 
 
 

Содержание TLU

Страница 1: ...ENGLISH ENGLISH ENGLISH ENGLISH Installation and User Manual version 1 04 TLU TLUANA 2014 30 EU EN55022 2010 EN61000 6 2 2005 EN61000 6 4 2007 SYSTEM IDENTIFICATION...

Страница 2: ...er No guarantee against misuse Batteries Laumas provides 1 year guarantee from the date of delivery note against material defects or battery manufacturing faults Disposal of Waste Equipment by Users i...

Страница 3: ...M CAPACITY 11 TARE WEIGHT ZERO SETTING 11 ZERO VALUE MANUAL ENTRY 11 REAL CALIBRATION WITH SAMPLE WEIGHTS 12 FILTER ON THE WEIGHT 13 ZERO PARAMETERS 13 RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANG...

Страница 4: ...SAMPLE WEIGHT 26 9 KEYPAD LOCK BLOCK THE ACCESS TO THE INSTRUMENT 27 10 KEYPAD UNLOCK 27 11 DISPLAY AND KEYPAD LOCK 27 12 CHECK SUM CALCULATION 27 MODBUS RTU PROTOCOL 28 FUNCTIONS SUPPORTED IN MODBUS...

Страница 5: ...e RC filters on the instrument driven solenoid valve and remote control switch coils Avoid inverters in the instrument panel if inevitable use special filters for the inverters and separate them with...

Страница 6: ...l shifting and tipping relating to shocks and vibration windy conditions seismic conditions in the installation setting stability of the support structure EARTHING THE WEIGHED STRUCTURE By means of a...

Страница 7: ...xcitation wire must be equal or similar to the one indicated in the load cell data sheet input resistance The insulation value between the shield and any other cell wire and between any other cell wir...

Страница 8: ...to zero set or display the net weight both values will be lost when the instrument is turned off Peak weight function Transmits the gross or net weight via optoisolated analog output 16 bit current 0...

Страница 9: ...150mA DIGITAL INPUTS N 2 optoisolated 5 24 VDC PNP SERIAL PORTS RS485 RS232 BAUD RATE 2400 4800 9600 19200 38400 115200 HUMIDITY non condensing 85 STORAGE TEMPERATURE 30 C 80 C WORKING TEMPERATURE 20...

Страница 10: ...nds of the network as described in the paragraph RS485 SERIAL CONNECTION The analog board ANALOG is present on TLUANA model WIRING DIAGRAM 13 1 2 3 4 5 6 7 8 9 1112 10 21 141516171819 20 mA V mA V COM...

Страница 11: ...vate the secondary LED function during weight display press and hold down MENU and keys at the same time press MENU immediately followed by KEY Short press Long press 3 s Into menus Tare resetting Can...

Страница 12: ...8 MENU MAP Into menus changes are applied right after pressing the ENTER key no further confirmation is required SETPOINT SYSTEM PARAMETERS 3 s TLUANA 3 s TLUANA...

Страница 13: ...necessary see section REAL CALIBRATION WITH SAMPLE WEIGHTS If the instrument HAS NOT BEEN CALIBRATED missing plant system identification tag proceed with calibration If load cells data are unknown fo...

Страница 14: ...s possible to set a value between 0 50000 and 7 00000 mV V Example of 4 cell system with sensitivity 2 00100 2 00150 2 00200 2 00250 enter 2 00175 calculated as 2 00100 2 00150 2 00200 2 00250 4 The d...

Страница 15: ...ompensate zero variations due to the presence of product residues Procedure Confirm the message Zero by pressing ENTER The weight value to be set to zero is displayed In this phase all of the LEDs are...

Страница 16: ...the system must read zero If this does not happen it means that there is a mechanical problem affecting the system linearity WARNING identify and correct any mechanical problems before repeating the...

Страница 17: ...fied again until an optimum result is achieved The filter enables to stabilise a weight as long as its variations are smaller than the corresponding response time It is necessary to set this filter ac...

Страница 18: ...al to 10 x SETTING UNITS OF MEASURE These are the available units of measure kilograms grams tons SEMI AUTOMATIC TARE NET GROSS THE SEMI AUTOMATIC TARE OPERATION IS LOST UPON INSTRUMENT POWER OFF To p...

Страница 19: ...eset tare value is set to zero The NET LED is turned off when the gross weight is displayed once again While the net weight is displayed keep pressed to display the gross weight When the key is releas...

Страница 20: ...the parameter to 8000 analog output correction to zero if necessary adjust the analog output allowing the PLC to indicate 0 The sign can be set for the last digit on the left E g if I use a 4 20 mA ou...

Страница 21: ...s set continuous weight transmission protocol to RIP6100 RIP675 RIP6125C series remote displays the remote display shows the net weight or gross weight according to its settings set continuous weight...

Страница 22: ...J2 J1 33 0 VDC 0 VDC 33 0 VDC 33 If the RS485 network exceeds 100 metres in length or baud rate over 9600 are used close the two jumpers called RS 485 termination to activate two 120 ohm terminating...

Страница 23: ...INGLE instrument to be used in a ONE WAY mode TEST Input Test ensure that for each open input is displayed is displayed when the input is closed Output Test setting ensure that the corresponding outpu...

Страница 24: ...set in this parameter from 0 to max 9 9 s default 0 delay time between the moment when the weight exceeds the set point and the moment when the relay opens for setpoint 1 only the possible values are...

Страница 25: ...les a number of decimals is selected for full scale which exceeds the instrument s display potential value above the maximum setting value the weight value set in sample weight verification does not m...

Страница 26: ...n and excluded The checksum value is obtained from the calculation of XOR or exclusive of the 8 bit ASCII codes of the characters considered This obtains a character expressed in hexadecimals with two...

Страница 27: ...section CHECK SUM CALCULATION CR 1 character for string end 13 ASCII 1 character of separation 92 ASCII 1 SETPOINT PROGRAMMING 1 1 SETTING SETPOINT VALUES CURRENTLY IN USE The PC transmits the followi...

Страница 28: ...alue is under 99999 the minus sign is sent alternated with the most significant figure Error messages In case of an instrument alarm for exceeding 110 of the full scale or 9 divisions above the value...

Страница 29: ...ng aaDckckCR Possible instrument responses correct reception aaxy ckckCR incorrect reception aa ckckCR where x number of decimals y value of division The y field can have the following values 3 for di...

Страница 30: ...w the operator to check the calibration obtained by using sample weights and correct automatically any change between the displayed value and the correct one Load onto the weighing system a sample wei...

Страница 31: ...s The check digit is calculated by executing the operation of XOR exclusive or of 8 bit ASCII codes of only the string underlined The procedure to perform the calculation of check sum is the following...

Страница 32: ...te useless operations on said locations The instrument in any case makes sure that no writing occurs if the value to be memorised is equal to the value in memory The numerical data found below are exp...

Страница 33: ...ber of Modbus registers to write beginning from the address no 1 No byte Number of bytes of the following data FUNCTION 16 Preset multiple registers WRITE MULTIPLE DI REGISTERS QUERY Address Function...

Страница 34: ...is instrument can manage a maximum of 32 registers read and written in a single query or response R the register may only be read W the register may only be written R W the register may be both read a...

Страница 35: ...lues the analog output zero and full scale values are saved in RAM they are lost upon shut down To save them permanently in the EEPROM so that they remain upon re start it is necessary to send the com...

Страница 36: ...less than 999999 Bit 5 Net weight beyond 999999 or less than 999999 Bit 6 Bit 7 Gross weight negative sign Bit 8 Net weight negative sign Bit 9 Bit 10 Net display mode Bit 11 Weight stability Bit 12 W...

Страница 37: ...instrument Least significant byte L Byte Most significant byte H Byte Division value Divisor Decimals Unit of measure value Unit of measure description Utilisation of the Coefficient with the differe...

Страница 38: ...y see section SEMI AUTOMATIC TARE NET GROSS 24 8 SEMI AUTOMATIC ZERO 99 Save data in EEPROM 9 GROSS display see section SEMI AUTOMATIC TARE NET GROSS 100 Zero setting for calibration see section TARE...

Страница 39: ...L h10 Number of registers to send H h00 Number of registers H h00 Number of registers to send L h02 Number of registers L h02 Byte Count h04 CRC16 H h40 Datum 1 H h00 CRC16 L h0D Datum 1 L h00 Datum...

Страница 40: ...1 h03 h00 h07 h00 h04 hF5 hC8 The instrument will respond with the string h01 h03 h08 h00 h00 hF hA0 h00 h00 h0B hB8 h12 h73 Query field name hex Response field name hex Instrument Address h01 Instrum...

Страница 41: ...indicate that this menu item is unlocked TEMPORARY MENU UNLOCKING press and simultaneously for 3 seconds it is now possible to enter and modify all menus including those which are locked By returning...

Страница 42: ...erased If you do not have a specific manual for the newly set program you can request it to technical assistance KEYPAD OR DISPLAY LOCKING Press ESC immediately followed by hold them down for about 5...

Страница 43: ...henden Normen bereinstimmt F D claration de conformit Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s CZ Prohl...

Отзывы: