- 11 -
MAXIMUM CAPACITY
: Maximum displayable weight (from 0 to max full scale; default: 0). When the weight exceeds
this value by 9 divisions, the display shows
. To disable this function, set 0.
TARE WEIGHT ZERO SETTING
This menu may also be accessed directly from the weight display, holding down the
0
key for 3 seconds.
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
-
Confirm the message
(Zero) by pressing ENTER.
-
The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing.
-
Confirming once again, the weight is set to zero (the value is stored to the permanent memory).
-
Press
▲
to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
ZERO VALUE MANUAL ENTRY
WARNING:
Perform this procedure only if it’s not possible to reset the weighed structure tare, for
example because it contains product that can not be unloaded.
Set in this parameter the estimated zero value (from 0 to max 999999; default: 0).