- 21 -
EVENTS LOG
The instrument can store up to 50 events; the oldest records are overwritten.
-
: it displays the last 50 events stored, starting from the most recent one:
: zero-setting from the calibration menu, press ENTER to display the value set to zero.
: theoretical calibration, press ENTER to display the full scale set.
: real calibration, press ENTER to display the sample weight used.
: tare setting via the keypad, press ENTER to display the set value.
: load distribution error, press ENTER to display the weight value that triggered the
alarm, press ENTER again to display the difference between the load percentage and the
stored value.
: weight alarm, press ENTER to display the alarm type.
: load distribution deletion.
: equalization.
: modification or deletion of the manual selection of active channels.
-
:
it deletes stored events; confirmation is requested (
),press ENTER to proceed
or press ESC to cancel.
-
:
it prints all events.
INFO MENU
: the identification data of the instrument are displayed.
-
: instrument model
-
: software code
-
: program type
-
: software version
-
: serial number
: the main legally relevant parameters are displayed.
-
: approval status of the instrument
-
: number of qualified accesses
-
: system full scale
-
: rated sensitivity of the load cells
-
: scale verification division for single interval instruments
-
: scale verification division for the first range of multi-interval or multiple range
instruments
-
: scale verification division for the second range of multi-interval or multiple range
instruments
-
: scale verification division for the third range of multi-interval or multiple range
instruments
+
+