82 / 137
Proline Kryomats
When a setpoint step change is specified, it may be that the optimum control would set a bath temperature, which
might significantly exceed the temperature desired on the external vessel. There is a correction limitation, which
specifies the maximum permissible deviation between the temperature on the external load and the heat transfer liquid
temperature.
Command
Control Parameters
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to:
Control
Control
Parameters
Control Parameters
.
The adjacent window appears.
Parameter endings:
e = Master controller, f = Slave controller.
Where applicable change parameters marked
with (auto) to manual input with
Tv
manual/auto
.
Select the parameters to be changed with
and confirm with
.
Then change the value in the following settings
window and confirm with
.
Correction limitation
see introduction.
7.11.2.1
Steps for setting the control parameters for external control
1.
Activate external control
2.
Setting the slave controller:
2.1.
Set parameters to
auto
; Check for thermostat type and change when necessary (RP....)
Choose heat transfer liquid with low viscosity and high thermal capacity.
Ranking: water, ethanol, water-glycol, oil, Fluorinert®.
Set pump level as high as possible,
make bath circulation strong and fast,
choose hose length as short as possible, i.e. 2 x 1m,
choose hose cross section as large as possible, i.e. ½ inch,
throughput through the external load as large as possible.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Kpe
0,50
Tne
100
Tve (auto) 83
Tde (auto)
8,3
Xpf
4,0
Prop_E(a) 30