36 / 137
Proline Kryomats
The external application must not have a hydraulically blocking effect.
If cross-sectional area of tube is too low
temperature gradient between bath and external
load due to low flow rate.
Always ensure the largest possible passages in the external circuit.
When tightening the union nuts on the pump nipple AF 19, use a wrench AF 14 to counter
the tightening torque (see figure).
If external control is to be used, provide a Pt100 temperature probe in the external load (
With loads at a higher position and with stationary pump and ingress of air into the
thermostatic circuit, the external volume can drain away, even with closed circuits
Risk of
thermostat overflowing!
Secure hoses against slippage with hose clips.
Unused pump connectors must be closed off.
Содержание RP 4050 C
Страница 130: ...130 137 Proline Kryomats 03 2022 12 Declaration of conformity and product returns declaration...
Страница 131: ...03 2022 Proline Kryomats 131 137...
Страница 134: ...134 137 Proline Kryomats 03 2022...
Страница 135: ......