5 Before starting up
5.1
Installing the interface module
The interface module is connected to an internal LiBus ribbon cable and
inserted into a vacant module slot. The number and arrangement of the
module slots vary depending on the device. The module slots are protected
by a cover that is screwed onto the casing or attached to the slot opening.
WARNING!
Touching live parts
Electric shock
Disconnect the device from the power supply before
starting any installation work.
Always observe safety measures against electrostatic dis-
charge.
The module installation description essentially applies to all LAUDA
constant temperature equipment; the example diagrams here show
the installation of an analog module in constant temperature equip-
ment from the Variocool product line.
Please note that an interface module with a small cover can only
be installed in a low module slot. The fitted cover must cover the
opening on the module slot completely.
You will require two M3 x 10 screws and a suitable screwdriver to
secure the interface module.
Please observe the following installation sequence:
1.
Turn off the constant temperature equipment and pull out the mains
plug.
2.
If necessary, remove the screws from the cover on the relevant
module slot. If necessary, use a slotted screwdriver to prise off the
cover.
Fig. 2: Removing the cover (schematic dia-
gram)
V1
Interface module LRZ 930
11 / 38
Содержание LRZ 930
Страница 38: ...Terminal program 31 U Update 14 W Warning 33 Warranty 6 V1 Interface module LRZ 930 38 38 ...
Страница 39: ......