6 Commissioning
6.1
Contact assignment of Ethernet interface
The Ethernet interface is equipped with standard type RJ45 sockets (8P8C
modular plugs according to CFR Part 68). Standard Ethernet cables that
correspond to category CAT5e or higher (8P8C assignment with twisted
pairs) must be used for the connection.
Table 1: RJ45 contact assignment
Contact
Signal 10Base-T / 100Base-TX
1
Tx+
2
Tx-
3
Rx+
4
–
5
–
6
Rx-
7
–
8
–
6.2 Software update
Older software installed on constant temperature equipment may have to be
updated for the new interface to work.
1.
Switch on the constant temperature equipment after installing the
new interface.
2.
Check whether a software warning appears on the display:
Warning
SW too old
: Please contact LAUDA Service, see
Chapter 1.6 “Contact LAUDA” on page 6.
No software warning: Operate the constant temperature equip-
ment as normal.
Fig. 7: RJ45 contacts on socket / plug
V2R4
Interface module LRZ 921