Control parameters
The control parameters have been optimized at the factory for operation as a bath thermostat (with water as the heat
transfer liquid) with internal control. The standard parameters are already set as default also for the thermostatic control
of external applications with external control.
Depending on the application, the configuration can be adapted from case to case as required. Also the thermal capacity
and the viscosity of the heat transfer liquid affect the control behavior.
Note: Only change the control parameters if you have adequate knowledge of control techniques.
17.1
Internal control variable (internal temperature sensor)
If you have not connected any temperature sensor, read further here. For activated external control read (
The control corresponds to the set-point temperature with the current bath temperature and calculates the set value for
heating or cooling.
These control parameters can be set:
Description
Short form
Unit
Proportional range
Xp
K
Reset time
Tn
s
Derivative time
Tv
s
Damping time
Td
s
If "
Tv manual/auto
" is set to "auto" (automatic),
Tv
and
Td
cannot be changed. They are in this case derived from
Tn
with fixed factors.
Consider the effect of the temperature limits Tih and Til (
7.4.5) on the control.
Содержание ECO GOLD
Страница 1: ...Operation manual ECO Gold ...
Страница 2: ......
Страница 4: ......
Страница 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Страница 82: ...82 137 ECO Gold 10 2019 i ...
Страница 83: ...10 2019 i ECO Gold 83 137 ...
Страница 84: ...84 137 ECO Gold 10 2019 i ...
Страница 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Страница 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Страница 130: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......