ECO Gold
101 / 137
The tolerance entry can have a large effect with external
bath control. The adjacent graph of the edited trace shows
the possible run-on of the actual temperature in the bath
vessel (continuous line) for the set-point temperature of
the programmer (highlighted in gray).
Note:
The tolerance field facilitates exact conformance
to the dwell time at a specified temperature. The
following segment is only processed when the ac-
tual temperature reaches the tolerance band
,
so that for example the ramp of segment 2 is only
started delayed by
.
A tolerance range which is too tight can however also cause undesired delays. In particular with external control
the range should not be chosen too tightly. In Segment 5 a larger tolerance has been entered, so that the de-
sired time of ten minutes is maintained even with settling action
.
Only flat (slow) ramps should be programmed where necessary with a tolerance range. Steep ramps which lie
close to the maximum possible heating or cooling rates of the thermostat may be severely delayed by a toler-
ance range that is too tight (here in Segment 2)
.
Note: No time specification is possible in the start segment (No. 1). The temperature of the first segment is attained as
quickly as possible in order to switch to segment 2 after reaching the set tolerance.
Содержание ECO GOLD
Страница 1: ...Operation manual ECO Gold ...
Страница 2: ......
Страница 4: ......
Страница 81: ...10 2019 i ECO Gold 81 137 W Warnings 55 92 X Xp 106 Xpf 108 ...
Страница 82: ...82 137 ECO Gold 10 2019 i ...
Страница 83: ...10 2019 i ECO Gold 83 137 ...
Страница 84: ...84 137 ECO Gold 10 2019 i ...
Страница 85: ...10 2019 i ECO Gold 85 137 Appendix with settings ...
Страница 121: ...10 2019 i ECO Gold 121 137 Click on the tab with the mouse and then on the Device manager ...
Страница 130: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......