8 Operator’s Manual
1.4 STERILIZATION
Laserscope recommends that GreenLight HPS fibers be used in only one
surgical procedure. The GreenLight HPS fiber is a sterile, single-use,
disposable device and should not be re-used or re-sterilized. After the
completion of the procedure, dispose of the GreenLight HPS fiber following
the standard hospital protocol for bio-hazardous material. The GreenLight
HPS fiber is ethylene oxide sterilized and is not designed or certified for
multiple use.
Do not re-sterilize. Do not reuse.
ADDITIONAL INFORMATION ABOUT REUSING THE GreenLight
HPS FIBER
If you need more information on this subject, we recommend the AAMI
Technology Assessment Report on the Re-Use of Disposables. It can be
ordered from the Association for the Advancement of Medical
Instrumentation at: (703) 525-4890 (Part Number: TAR6-058). Other
questions can be directed to the Laserscope Customer Response Center at
(800) 356-7600 or (408) 943-0636.
STORAGE
The GreenLight HPS fiber should be stored under ambient conditions.
WARRANTY
The warranty for the GreenLight fiber optic covers only defects in the
material and workmanship. Should you need to return this product, call the
Laserscope Customer Response Center for “Return Authorization”.
Instructions for returning a product to Laserscope are detailed in the
Accessory Return/Repair Policy included with each device.
¡
WARNING: Never secure a fiber optic using a clamp as doing so
can result in an unsafe use condition. Use of a clamp can result
in the fiber optic being bent at sharp angles or fiber damage
which, in turn, can also result in an unsafe use condition. The
fiber optic can break causing a burn in the protective jacket and
the release of laser energy. If undetected, this condition can
result in a burn or ignition of flammable materials.
When using the fiber with the laser, the fiber device connector
should never be allowed to touch the floor or any wet or non-
sterile surface. If the fiber connector does come into contact
with a surface other than the laser device port, dispose of the
fiber in accordance with hospital safety regulations.