BL57-NE
3-12
(E)
5
Remove the stopper from the detector head.
Fig. 3-16
6
Move the scale, and make sure that the amplitudes A and B (see Fig. 3-17) of the Lissajous’ figure are
within the range of the specified values over the entire length.
<Specified values>
A/B signal output type ..... Output format A : Amplitude 0.9 Vp-p to 1.8 Vp-p
Output format F : Amplitude 0.7 Vp-p to 1.3 Vp-p
Output format G : Amplitude 0.8 Vp-p to 1.2 Vp-p
Analog output type ........... Output format H : Amplitude 0.6 Vp-p to 1.1 Vp-p
For output format A, the center of the Lissajous’ figure is 0 V. For output format F, G, and H, the center
of the Lissajous’ figure is 2.5 V.
If the output of 1 Vp-p cannot be obtained even through repeated adjustment, recheck the mounting
surface accuracy of the scale and detector head (refer to pages 3-3 to 3-4), and check again that there is
no dirt on the scale’s surface.
Fig. 3-17
Lissajous
B
A
Stopper
Do not move the machine device with
the stopper attached to detector
head because this may result in
breakage of the detector head or the
scale.