Section 4 - System Functions
Page 17
Aligning the Tilt Sensor
In the rare event that the unit suffers a severe drop shock, refer to the
instructions below to align the tilt sensor.
1. Press the MODE button until the external LCD displays the SYS
Mode Indicator.
2. Press the FWD or BACK button until the CAL Function Indicator is
displayed and the message “rEF 1” appears in the main display.
3. Position the RD 1000 on a flat, relatively level surface as shown in
Figure #10(A).
¶
Do not lift or raise the unit until the message “dOnE” is
displayed in step #6 below.
4. Press the ENTER button to store the inclination measurement. The
message “rEF 2” appears in the main display.
5. Slowly rotate the RD 1000 180 degrees as shown in Figure #10(B).
6. Press the ENTER button to store the inclination measurement and
make the zero offset correction. The message “dOnE” appears in the
main display before the message “ref 1” is displayed.
¶
Press the FWD button to move to the next System option.
¶
Press the BACK button to move to the previous System option.
¶
Press the MODE button to select a measurement mode.
L
When using an LTI laser range finder with your RD 1000, you should
verify that you have selected the same measurement units in both
instruments. If the RD 1000 and the laser are not using the same
measurement units, the RD 1000 will convert the measurement data
received from the laser. For example, if the RD 1000 is using
Imperial
units
, and the laser is using
Metric units
, the RD 1000 will convert the
measurement data received from the laser to
Imperial units
.
Figure #10