EL-SIE-2+

High Accuracy Temperature and Humidity 

USB Data Logger, EasyLog Cloud Compatible

When tracking changes in ambient conditions, the response time of the humidity sensor in your data logger is approximately 20 minutes 
to reach 90% of the reading. However, if you are measuring step changes in humidity (for example if calibrating the product) it is advised 
that you leave the unit for up to four hours to ensure that it has enough time to settle at the new level.

It is worth remembering that the value of relative humidity is of course sensitive to temperature variation. As an example, at a relative 
humidity of ~90%RH at ambient temperature, a variation in temperature of 1°C will result in a change of up to -5%RH. Therefore when 
comparing multiple devices or calibrating them, any temperature variations must be considered.

The humidity measuring element in the humidity data loggers can be contaminated through exposure to a variety of compounds. These 
products should not be kept in proximity to volatile chemicals such as solvents and other organic compounds. Generally speaking, if a 
material or compound emits a strong odour you should not keep your humidity data logger in close proximity to it. If you would like more 
information, please contact your local Lascar Electronics office.

High levels of pollutants may cause permanent damage to the internal sensor.

BATTERY INFORMATION

We recommend that you replace the batteries annually, or prior to logging critical data. Use only AAA 1.5V alkaline batteries. Before 
replacing the batteries, disconnect the logger from your computer.

The logger does not lose its stored data readings when the batteries are discharged or replaced. However, the logging process will stop 
and will not resume until the batteries are replaced and the logger is connected to your computer and the logger is started.

Note that while the logger is plugged into a computer it draws power from the USB port instead of the batteries, which can raise the 
temperature of the logger slightly. It will return to normal shortly after disconnection.

Содержание EL-SIE-2+

Страница 1: ...s over 1 000 000 readings As per spec table No software to install configure using your normal web browser Use with a PC or Mac Display shows current maximum and minimum readings Status and alarm indicators On board alarm sounder Data can be uploaded to the EasyLog Cloud ACCESSORIES BAT 1V5 AAA 2 x AAA 1 5V alkaline batteries CABLE USB 3 1 C 1M USB A to USB C Cable INCLUDED IN THE BOX BAT 1V5 AAA ...

Страница 2: ...arameters to be configured Logger and channel names Measurement units Logging rate and start mode Up to 16 separate alarms high low pre alarm cumulative with thresholds delay and hold Display and LED indicator modes Once the logger is running you can plug it back into your computer and see the latest data device status and the event log You can also chose to stop the logger and change the configur...

Страница 3: ...e is reached Low Battery The batteries are low and should be changed when possible Memory Full The memory is full and logging has stopped Memory 90 The memory is 90 full and data should be downloaded when possible Calibration Pending The calibration will expire in 30 days on the date shown can be DD MM YYYY or MM DD YYYY Calibration Expired The calibration has expired on the date shown can be DD M...

Страница 4: ...creen Button Press Function Event Recorded USB n a n a n a n a Push to Start Any Long Start logging n a Triggered Start n a n a n a n a Delayed Start n a n a n a n a Channel Current reading Top Short Move to next channel or STOP LOG n a Long Clear alarm hold for all channels Clear Held Alarms Bottom Short Show Min reading for this channel n a Long Mute alarm sounder Mute Alarm Channel Min reading ...

Страница 5: ...hows best performance when operated within the recommended normal temperature and humidity range of 5 to 55 C and 20 to 80 RH respectively Long term exposure to conditions outside the normal range especially at high humidity may temporarily offset the RH signal e g 3 RH after 60h kept at 80 RH After returning into the normal temperature and humidity range the sensor will slowly come back within th...

Страница 6: ...d through exposure to a variety of compounds These products should not be kept in proximity to volatile chemicals such as solvents and other organic compounds Generally speaking if a material or compound emits a strong odour you should not keep your humidity data logger in close proximity to it If you would like more information please contact your local Lascar Electronics office High levels of po...

Отзывы: