SPECIFICATIONS 

Calibration Traceability 

Larson Electronics

 solar radiation meters are calibrated through side-by-side comparison to the mean of four  

transfer standard pyranometers (shortwave radiation reference) under high intensity discharge metal halide 
lamps. The transfer standard pyranometers are calibrated through side-by-side comparison to the mean of at least 
two ISO-classified reference pyranometers under sunlight (clear sky conditions) in Logan, Utah. Each of four ISO-
classified reference pyranometers are recalibrated on an alternating year schedule (two instruments each year) at 
the National Renewable Energy Laboratory (NREL) in Golden, Colorado. NREL reference standards are calibrated to 
the World Radiometric Reference (WRR) in Davos, Switzerland. 

ISO 9060:2018 

Class C (previously known as second class

Calibration Uncertainty 

± 5 % (see Calibration Traceability below) 

Measurement 
Repeatability 

Less than 1 % 

Long-term Drift 
(Non-stability) 

Less than 2 % per year 

Non-linearity 

Less than 1 % (up to 1750 W m

-2

Response Time 

Less than 1 ms 

Field of View 

180º 

Spectral Range 

360 to 1120 nm (wavelengths where response is 10 % of maximum; 

see Spectral Response below) 

Directional (Cosine) 
Response 

± 5 % at 75º zenith angle (see Cosine Response below) 

Temperature Response 

0.04 ± 0.04 % per C (see Temperature Response below) 

Operating Environment 

0 to 50 C; less than 90 % non-condensing relative humidity up to 30 C; less than 

70 % non-condensing relative humidity from 30 to 50 C; separate sensors can be 

submerged in water up to depths of 30 m 

Meter Dimensions 

126 mm length; 70 mm width; 24 mm height; 

Sensor Dimensions 

24 mm diameter; 33 mm height 

Mass 

150 g 

180 g 

Cable 

2 m of two conductor, shielded, twisted-pair wire; TPR jacket (high water 

resistance, high UV stability, flexibility in cold conditions) 

Larson Electronics, LLC

 

  Phone: (800) 369-6671 

 Fax: (903) 498-3364 

  www.larsonelectronics.com 

Содержание XLE-HHLM-VISNIR-360NM.1200NM-V1

Страница 1: ...OWNER S MANUAL SOLAR RADIATION METER Part XLE HHLM VISNIR 360NM 1200NM V1 Larson Electronics LLC Phone 800 369 6671 Fax 903 498 3364 www larsonelectronics com...

Страница 2: ...rnational Organization of Standardization ISO second class and first class classifications but because of limited spectral sensitivity they do not meet the spectral specification necessary for WMO or...

Страница 3: ...rtainty 5 see Calibration Traceability below Measurement Repeatability Less than 1 Long term Drift Non stability Less than 2 per year Non linearity Less than 1 up to 1750 W m 2 Response Time Less than...

Страница 4: ...a for the photodiode were obtained from the manufacturer Mean temperature response of four Larson silicon cell pyranometers Temperature response measurements were made at approximately 10 C intervals...

Страница 5: ...rado Cosine response was calculated as the relative difference of pyranometer sensitivity at each solar zenith angle to sensitivity at 45 solar zenith angle The blue symbols are AM measurements the re...

Страница 6: ...ogged measurements on the LCD display Logging To choose between manual or automatic logging push the mode button once and use the up down buttons to make the appropriate selection SMPL or LOG Once the...

Страница 7: ...fers in clear sky and cloudy conditions thus measurements in cloudy conditions result in spectral error because sensors are calibrated in clear sky conditions Spectral response of Larson series pyrano...

Страница 8: ...m Agricultural Soil 12 1 Reflected from Forest Soil 4 1 Reflected from Desert Soil 3 0 Reflected from Water 6 6 Reflected from Ice 0 3 Reflected from Snow 13 7 Spectral error for Larson series pyranom...

Страница 9: ...acy the sky must be completely clear as reflected radiation from clouds causes incoming radiation to increase above the value predicted by the clear sky calculator Measured values of total shortwave r...

Страница 10: ...nt draw This is necessary to maintain the logged measurements in memory Therefore it is recommended to remove the battery when storing the meter for many months at a time in order to preserve battery...

Страница 11: ...PU voltage below minimum Fix replace CR2320 battery and perform master reset Larson Electronics LLC Phone 800 369 6671 Fax 903 498 3364 www larsonelectronics com THESE INSTRUCTIONS MAY NOT COVER ALL D...

Отзывы: