PRM2103 Reference Manual
Performing Calibration Checks
B-12
Figure B-16 shows the resulting Calibration Check message and the
prompt to store the Standard Calibration Check Level.
FIGURE B-15
.Calibration Check
TAKE NOTE
If you click Yes to
this prompt and store a new
Standard Calibration Check
Level, all subsequent
calibration checks in the
Calibration Check History will
measure deviations from this
check.
FIGURE B-16
.Calibration Check Message and Prompt
Click No to simply check the value and have it stored in the Calibration
Check History. Click Yes to store a new Standard Calibration Check Level.
After clicking either option, the Calibration Check message indicates that
the calibration has succeeded, as shown in Figure B-17.
FIGURE B-17
Calibration Check Succeeded
B.4.5 Automatic Calibration Checks in SLM Utility-G3 Software
LEARN MORE
For more
information on setting up
calibration checks or viewing
calibration data in the SLM
Utility-G3 software, refer to
SLM Utility-G3 Software
Manual.
To schedule automatic calibration checks in SLM Utility-G3 software,
follow these steps:
Step 1
Click the Measurement Control tab after clicking the Settings
tab.
Step 2
Select the Enable Measurement History option, if desired.
Содержание PRM2103
Страница 1: ...Model PRM2103 Outdoor Microphone Preamplifier ...
Страница 2: ...I2103 01 Rev C Larson Davis PRM2103 Preamplifier Reference Manual ...
Страница 34: ...AEC206 Reference Manual Electrical A 4 FIGURE A 2 1 3 Octave Self Generated Noise ...
Страница 37: ...AEC206 Reference Manual Physical A 7 Effect of Humidity FIGURE A 5 Relative SPL versus Humidity ...
Страница 38: ...AEC206 Reference Manual Physical A 8 Humidity Endurance FIGURE A 6 Relative SPL versus Humidity Endurance ...
Страница 91: ...PRM2103 Reference Manual Upgrading Firmware with SLM Utility G3 B 16 ...