PRM2103 Reference Manual
Performing Calibration Checks
3-10
3.4 Performing Calibration Checks
TAKE NOTE
The calibration
check level reported by the
SoundAdvisor is the
broadband level.
The PRM2103 calibration check function produces five discrete tones
simultaneously. Calibration checks can be performed automatically at
specified times, or at any time manually, either on the SoundAdvisor or G4
LD Utility.
3.4.1 Manual Calibration Checks
To perform manual calibration on the SoundAdvisor, follow these steps:
TAKE NOTE
The standard check
level is set by performing a
manual Cal Check directly after
performing an acoustic
calibration.
Step 1
After performing an acoustic calibration, navigate
Tools
Calibration
Cal Check
tab.
FIGURE 3-13
Cal. Check Tab
The Standard Cal. Level field holds the last saved cal check value. The
default value is 94.00 dB.
You can manually change the dB level by navigating to the field and
changing the number value.
After performing the calibration check, the meter displays a message
prompting you to save a new Standard Cal. Level, if needed, as shown in
Figure 3-14.
Содержание PRM2103
Страница 1: ...Model PRM2103 Outdoor Microphone Preamplifier ...
Страница 2: ...I2103 01 Rev C Larson Davis PRM2103 Preamplifier Reference Manual ...
Страница 34: ...AEC206 Reference Manual Electrical A 4 FIGURE A 2 1 3 Octave Self Generated Noise ...
Страница 37: ...AEC206 Reference Manual Physical A 7 Effect of Humidity FIGURE A 5 Relative SPL versus Humidity ...
Страница 38: ...AEC206 Reference Manual Physical A 8 Humidity Endurance FIGURE A 6 Relative SPL versus Humidity Endurance ...
Страница 91: ...PRM2103 Reference Manual Upgrading Firmware with SLM Utility G3 B 16 ...