PRM2103 Reference Manual
Performing Calibration Checks
B-8
FIGURE B-8
PRM2103-FF Data on the Instrument Status Tab
B.4 Performing Calibration Checks
TAKE NOTE
The calibration
check level reported by the
Model 831 is the broadband
level.
The PRM2103-FF calibration check function produces five discrete tones
simultaneously. Calibration checks can be performed automatically at
specified times, or at any time manually, either on the Model 831, with
SLM Utility-G4, or with SLM Utility-G3 software.
B.4.1 Manual Calibration Checks on the Model 831
To perform manual calibration on the Model 831, follow these steps:
Step 1
Press the
3
(TOOLS) key.
Step 2
Select the Calibration icon and press
5
.
Step 3
Navigate to the Cal. Check tab.
Step 4
Enable the Cal. Check button and press
5
.
Figure B-9 shows the Cal. Check tab on the Calibrate Tool.