Conductivity measurement
23
• Setting the TDS method
TDS is calculated from the measured conductivity value. The available calibration
methods are "Linear": KCl with factor adjustable from 0.40 to 1.00 (default: 0.50), "442":
Myron L 442 non-linear standard curve, "En": European environmental standard non-
linear curve, and "NaCl": non-linear salinity curve.
1. Press the
key to enter the setting mode.
2. Press the
keys to select "TDS" (total dissolved solids setting) and then
press the
key.
3. Press the
keys to select "TYPE" (TDS method setting) and then press the
key.
key
key
key
key