Conductivity measurement
21
• Setting the temperature conversion (Default: ON, 2.00%/°C)
The measured value of a sample that is not at 25°C can be converted to a value at the
selected temperature. To use the temperature conversion function correctly, temperature
coefficient (the rate of change per 1°C of the conductivity) must be set for each sample.
The setting of “ Setting the temperature display ” (page 16) is applied to the sample
temperature before the conversion.
1. Press the
key to enter the setting mode.
2. Press the
keys to select "COND" (conductivity setting) and then press the
key.
3. Press the
keys to select "TC" (temperature conversion setting) and then
press the
key.
4. Press the
keys to enter the temperature coefficient and then press the
key to confirm.
To return to the measurement mode, press the
key.
Note
・
The temperature coefficient varies by sample. Before using the temperature conversion
function, always check the temperature coefficient of the sample and set it in the
instrument.
・
When the temperature conversion function is used with automatic temperature
measurement (ATC), deviations may occur within the accuracy of the temperature sensor.
For more accurate measurement, set the temperature setting to manual temperature
display (MTC), and measure using a temperature controlled bath.