Page 12 of 43
Operating Manual
www.qedenv.com
SEM5000.002
© QED Environmental Systems Inc.
SEM5000 Surface Emissions Monitor
NOTE: Once logging has been turned “ON” or “OFF” it will remain in that mode until
changed by the user.
The “Log” function must be turned “ON” in order for the SEM5000 to automatically save
survey readings.
When the “Log” function is not activated, the word “OFF” will appear at the bottom of
the display screen.
When the “Log” function is activated, the percent symbol, (%), will be displayed at the
bottom of the screen. During the survey, this will indicate the percentage of available
memory that has been used.
•
Press the Selector dial and scroll to the “Select Save Interval” option and push the
Selector dial to display time save options
•
Scroll to select auto save time, (1 second to 10 seconds), and press again to confirm
your selection
NOTE: If you do not wish to utilize the Auto Save function, bypass the “Select Save
Interval” option and instead activate the “Manual Save” option.
This option will allow you to use the “Function” button to manually save readings at
desired intervals.
•
The instrument is now ready for field calibration