Page 11 of 43
Operating Manual
www.qedenv.com
SEM5000.002
© QED Environmental Systems Inc.
SEM5000 Surface Emissions Monitor
NOTE: The GPS module is a high-performance receiver with the ability to track up to 20
satellites.
Start up time for the GPS module can vary according to the strength of the actual signal
received.
The presence of high trees or buildings, aerial power lines or other obstructions will affect
the signal. Typically, in an open area, the GPS will fix in 1 to 3 minutes.
In the event of signal loss during the survey, the instrument will continue to log gas values
and will associate the to the last saved tracked GPS position until the signal is reacquired.
SEM5000 Set Up
Once GPS has been activated, survey parameters will need to be set up. Gas Exceedance
levels and data storage preferences must be selected.
•
Push the Selector dial to display the Option menu and scroll to the “Select Gas
500/250/100” option and press the Selector dial to display the gas concentration
menu
•
Scroll to choose 500, 250 or 100 parts per million and push the Selector dial to store
your choice
•
Push the Selector dial and scroll to the “Log” option
•
Press the selector dial again to choose “ON” or “OFF”
Icon
Description
Icon
Description
GPS active but waiting for
signal
GPS off
GPS active
Save location