Figure 11: Communication module slot

7.1 Installing the communication module

The communication modules are connected via a non-reactive

plug to enable installation or conversion at any time.

Note

Only ever touch the module on its plastic holder.

Note

To connect an external cable, open the cable sleeves so

that they tightly enclose the cable.

Proceed as follows to install a communication module:

If necessary, open the housing cover by loosening the screw.

For M-Bus and pulse output modules, also note the following

points:

Run the cable from the outside through the grommet.

Strip and connect the cable.

Secure the cable with the strain relief clamp.

Connect the cables to the module’s terminals.

Pull the cables out through the housing while inserting

the module.

First attach the contact surfaces of the module to the

module slot.

Gently push the module in.

Depending on the design of the housing, please also observe the

following points:

For IP 68 versions of the housing, tighten the cable gland.

For the IP 54 version of the housing, make sure that the

grommet is seated correctly.

Close the housing cover by tightening the screw (see

chapter Torque) and press the housing cover tightly into

place.

Note

You can find the technical details and data about the

communication modules in their documentation.

Note

No later than 60 seconds after installation, the meter

automatically 

 detects the inserted modules and is

ready for communication or pulse output.

7.2 Terminal blocks

2-pin or 4-pin terminals are used to connect the external lines to

the modules.

Stripping length 6 mm

Connection capability

rigid or flexible, 0.14 ... 1.5 mm²

flexible with wire end ferrule with plastic sleeve,

0.25 ... 1.5 mm²

flexible with wire end ferrule without plastic sleeve,

0.25 ... 1.0 mm²

Conductor sizes 26 ... 14 AWG

Recommended screwdriver:

0.6 × 3.5 mm

Tightening torque: 0.35 … 0.4 Nm

8 Non-detachable temperature sensors

Note

The cables must not be detached, shortened or extended

for factory-fitted, non-detachable temperature sensors.

Insert the temperature sensors into the immersion sleeves,

ball valves or T-pieces. The temperature sensor must sit on

the bottom of the thermowell.

Seal the temperature sensors to prevent manipulation.

9 Detachable temperature sensor

Note

If detachable temperature sensors are used, they must

have their own calibration or certificate of conformity.

Note

The maximum cable length of the temperature sensors is

10 m. An extension is not permitted.

If necessary, open the housing cover by loosening the screw.

Run the warm side temperature sensor cable from the

outside through the 4th grommet from the left and the cold

side temperature sensor of the through the 5th grommet

from the left.

Strip both cables as shown in the illustration.

© Gyr

Subject to change without prior notice

07.02.2020

6

Содержание T450

Страница 1: ...arantees Caution Comply with the requirements for circulation water CEN TR 16911 2016 Caution The meter is only suitable for circulation water in heating systems Caution The meter is not suitable for...

Страница 2: ...ion corresponds to the return and the cold side installation location to the supply Proceed as follows to install the meter 1 Compare the installation location with the symbol on the LCD or to determi...

Страница 3: ...Press button 2 briefly to complete the change The change can be checked using the symbols or on the LCD If necessary adjust the temperature sensors according to the installation requirements 4 2 Inst...

Страница 4: ...e temperature sensors are horizontal or vertical Figure 7 Recommended installation position of cooling meters and combined heat cooling meters 5 Calculator The ambient temperature of the calculator mu...

Страница 5: ...follows to change the battery If necessary open the battery compartment cover by loosening the screw Remove the old batteries from the battery compartment Insert the battery into the battery compartm...

Страница 6: ...on Note No later than 60 seconds after installation the meter automatically detects the inserted modules and is ready for communication or pulse output 7 2 Terminal blocks 2 pin or 4 pin terminals are...

Страница 7: ...be displayed without incorrect installation Error Incorrect flow direction negative Check whether the flow direction arrows on the volume meter correspond to the flow direction of the system If the di...

Страница 8: ...necessary or Error in internal memory Change the device The measurements of the meter may no longer be used for legally calibrated billing Error code Error Information for Service F2 F3 F5 or F6 pers...

Отзывы: