Landis+Gyr 750 Скачать руководство пользователя страница 6

 

Model 750 Gas Meter 

Installation, Commissioning and Maintenance Manual

 

 

Page 6 of 17 

© Gyr  

 

17-May-2016 

Due to continuous product development the details above are subject to change without notice. Whilst Gyr has taken all reasonable 

care in reviewing the information in this document, Gyr makes no representations in respect of, gives no guarantees or undertakings in 

relation to, and to the maximum extent permitted by law, excludes all warranties in relation to, the accuracy or completeness of the information 

in this document. Prospective users of this information should make their own enquiries as to its accuracy and reliability. 

accuracy  at  Qmax  relative  to  20%  of  Qmax).  The 
ability  to set timing  is  an  important  feature  of  the  750 
meter.  Like  a  well  tuned  engine  it  assures  both 
smooth operation and reduced wear. 

PRESSURE ABSORPTION 

 

Pressure  loss  is  made  of  two  components  pneumatic 
pressure loss (created by small passage ways and the 
movement  of  gas  through  the  meter)  and mechanical 
pressure loss a result of moving components. 

The  mean  pressure  absorption  of  the  Model  750  is 
125 Pascals (Pa) at a flow rate of 6.0 m

3

/hr of air and 

125 Pascals (Pa) at a flow rate of 7.5 m

3

/hr of natural 

gas (SG 0.62). 

ABSORPTION AND OSCILLATION 

Pressure

 

Absorption

  is  defined  as  the  average 

difference between the inlet and outlet pressure when 
the  meter  is  run  on  air  over  at  least  one  cycle  of  the 
meters  operation  measured  with  a  British  Slope 
Gauge. 

Pressure  Oscillation

  is  defined  as  the  difference 

between  the  highest  and  lowest  readings  of  a 
pressure  gauge  placed  at  the  outlet  of  the  meter 
measured with a British Slope Gauge. 

Although magnahelic and transducers can be used to 
determine  absorption  and  oscillation,  these  devices 
must  be  calibrated  to  the  reference  standard,  a  the 
British Slope Gauge as described in AG702. 

The  acceptance  criteria  for  pressure  absorption  is  no 
greater  than  125Pa  through  all  flow  rates  and  for 
oscillation 75Pa through all flow rates. 

 

METER ACCURACY TREND 

Trend  is  the  accuracy  at  100%  Qmax  minus  the 
accuracy  at  20%  badge  rate  shown  below.  Normally 
meters  are  set  with  a  negative  trend  rather  than  a 
positive  trend  due  to the  falling  off  in  accuracy  at low 
flow rates (pilot rate). 

 

The  750  high  resolution  adjustable  tangent  allows 
precision  fine  tuning  of  both  meter  Accuracy  and 
accuracy  trend  (accuracy  at  Qmax  relative  to  20%  of 
Qmax). The ability to set timing is an important feature 
of  the  meter.  Like  a  well  tuned  engine  it  assures 
smooth operation and reduced wear. 

This  is  not  available  with  fixed  tangent  meters  which 
use gear trains for pseudo accuracy correction without 
provision for timing adjustment. 

750's 

adjustment 

is 

self-locking. 

It 

ensures 

maintenance  of  factory  set  accuracy  during  transport 
and handling. 

ENDURANCE TESTING 

Ageing  life  tests  conducted  on  the  Model  750  meter 
have  provided  performance  results  that  demonstrate 
the  long  life  and  high  accuracy  for  in  excess  of  15 
million cycles. 

The  Model  750  meter  exceeds  the  endurance  testing 
requirements 

of 

the 

European 

Standard 

EN1359:1999,  American  National  Standard  ANSI 
B109.1, Australian Gas Association AG702, and OMIL 
Recommendation 

31. 

These 

standards 

and 

recommendations are summarised in Table 4 on page 
7. 

The endurance test results are shown on page 7 show 
the  equivalent  of  27  years  in  field  service,  assuming 
an  annual  consumption  1,100  cubic  metres  per 
annum. 

The flow rate during the test was continuous at flow 
rate of 6 m

3

/hr of natural gas (SG 0.62). 

 

Содержание 750

Страница 1: ... other database of these documents requires the written approval of Landis Gyr Due to continuous product development these details are subject to change without notice Model 750 Gas Meter Installation Commissioning and Maintenance Manual Revision 750002 Model 750 Installation Commissioning and Maintenance Manual 17th May 2016 Landis Gyr ...

Страница 2: ...______________________________________________________________8 STORAGE ________________________________________________________________________________8 TRANSPORTATION ________________________________________________________________________8 INSTALLATION ____________________________________________________________________________8 PUTTING INTO OPERATION ___________________________________________...

Страница 3: ...s that demonstrate the long life and high accuracy for in excess of 20 million cycles Exceeds the endurance testing requirements of both the European Standard EN1359 1999 and the Australian Gas Association AG702 which are in excess of the endurance testing requirements as specified in OMIL Recommendation 31 Pressure Absorption The slow revving operation 2 litres cycle the low friction valve assemb...

Страница 4: ...IMENSIONS MATERIAL OF CONSTRUCTION For more details refer to Document Number 750004 Model 750 Material Weight and Recyclability Table 1 Materials of Construction Material Housing Diecast Aluminium Centre Block PBT polybuylene terephthalate Grid Valves PF Phenoplastic Diaphragm Plates Pans Steel Rods Shafts Steel Valve Guide Steel Diaphragms Moulded PET and Elastomer Index window PC UV Stabilised P...

Страница 5: ...ation to pass initial testing however sustained accuracy over time comes down to the quality the design and construction of the meter Landis Gyr has always recognised that the cost of changing meters in the field coupled with repair costs is significant and a meter with a shorter field life also attracts other costs including uncertain revenue collection and hidden but real cost associated with a ...

Страница 6: ...ge Although magnahelic and transducers can be used to determine absorption and oscillation these devices must be calibrated to the reference standard a the British Slope Gauge as described in AG702 The acceptance criteria for pressure absorption is no greater than 125Pa through all flow rates and for oscillation 75Pa through all flow rates METER ACCURACY TREND Trend is the accuracy at 100 Qmax min...

Страница 7: ...mpleteness of the information in this document Prospective users of this information should make their own enquiries as to its accuracy and reliability INDEPENDENT ENDURANCE TESTING RESULTS 0 2 Qmax EN1359 1998 Endurance Test Results Conducted by Independent Accredited Laboratory DVGW Forschungstelle 3 2 1 0 1 2 3 0 5000 10000 15000 20000 25000 30000 Volume m3 5000h Qmax Percentage Error 13627419 ...

Страница 8: ...paired resulting in improper measurement and loss of revenue STORAGE Diaphragm meters should be stored upright in a clean dry place and fitted with dust caps on the inlet and outlet connections This will exclude debris that can be detrimental to the meter s internal components TRANSPORTATION Diaphragm meters should be transported upright be suitably restrained and properly cushioned to prevent the...

Страница 9: ...eam of the meter to be sure that all connections are made up and tight or that the downstream valve if there is one is closed CARE must be taken to prevent excessive flows and high pressure drops across the meter If CARE is not exercised when pressurising the meter over speeding of the meters internal mechanism may result in internal damage to the meter To prevent over speeding the pressure drop a...

Страница 10: ...phide and water Therefore the service life will be dramatically affected accordingly WARNING PURE OXYGEN All regulators and meters supplied by Landis Gyr are NOT SUITABLE for pressure regulating or measuring PURE OXYGEN Special meters and regulators made from specific materials are required to prevent heat build up as a result of friction etc SERVICE LIFE INTERNAL CORROSION Service life varies acc...

Страница 11: ...ning and closing of the reed switch All meters are fitted with a magnet in the last index wheel and hence each meter is pulse retro fittable while in service by use an External Pulse Module without the need to interfere with the metrology of the meter or the meter seals The pulse output is 1 pulse per 10 litres No other pulse output values are available Picture right shows the Model 750 Meter with...

Страница 12: ... 750085 PULSE OUTPUT INSTALLATION INSTRUCTIONS IN Z41 Assembly 1 Push connect pulse cable into Pulse Module Housing 1 IN Z41 Security Sealing Pin Wire 1 Insert sealing Pin 6 into Pulse Module Housing 1 2 Seal using wire length 9 and Lead Seal 10 if desired IN Z41 Installation to Index 1 Attach the pulse module to the meter index 2 Insert Copper Ring 3 and Lead Pin 2 3 Squash Lead Pin 2 using plier...

Страница 13: ...ed Plastic Seal Z4 Z5 3 Reqd 32446710 750013 X Red Plastic Seal NSW EL 3 Reqd 750014 X Red Plastic Seal 750 NZ 3 Reqd 750101 Change Gear for gas meter 3747000 750102 Change Gear for index 3747500 750020 Index Front Facing Complete 750021 Index Screw ISO 7045 M4x8 4 8 H Zn galv 4 Reqd 750022 Index Cover hydrophilic coated 7502XX X Index Plate Printed Customer Specific 750023 index plate source part...

Страница 14: ... o ring 34x2 5 NBR 70 Shore A 3110496 750070 X Meter Assembled w o Index or Badge c w Non Return 750065 Outlet Pipe Complete c w Non Return Valve 750064 Non Return Assembly 750061 Outlet Pipe Moulded 32319623 750062 Top o ring 27x2 NBR 70 Shore A 3110001 750063 Bottom o ring 34x2 5 NBR 70 Shore A 3110496 750031 O ring Top Bottom 3 5x190 NBR 70 Shore A 750066 Replaced by 750031 was 4x190mm 610019 M...

Страница 15: ...lation to the accuracy or completeness of the information in this document Prospective users of this information should make their own enquiries as to its accuracy and reliability Miscellaneous L0331 Grease Kluber Isoflex Topas NB5051 L0332 Oil Kluber Constant OY32 A0339 Pressure Test Nipple Sealant Accessories BKNut M5 Nyloc Nut 15006 M5x12 Screw Molybond Coated 750080 Box for Four 4 Meters 75008...

Страница 16: ...ng the information in this document Landis Gyr makes no representations in respect of gives no guarantees or undertakings in relation to and to the maximum extent permitted by law excludes all warranties in relation to the accuracy or completeness of the information in this document Prospective users of this information should make their own enquiries as to its accuracy and reliability ASSEMBLY DR...

Страница 17: ...able care in reviewing the information in this document Landis Gyr makes no representations in respect of gives no guarantees or undertakings in relation to and to the maximum extent permitted by law excludes all warranties in relation to the accuracy or completeness of the information in this document Prospective users of this information should make their own enquiries as to its accuracy and rel...

Отзывы: