Correcting Faults
Help:
If R
ki
-CO for LT3-F over 50
, and/or for LT3 over 100
, then
- Probe too cold.
- Internal resistance regulation not active or incorrect setpoint value.
- The probe is stuck in a cold spot and the heating capacity does not
suffice.
Look for a new installation position.
- Heating defective and probe cold.
- Sensor breakage.
If R
ki
-CO under 10
, then probe too hot.
- Internal resistance regulation not active or incorrect setpoint value.
The optimum operating point (setpoint value) lies at 20
007 (HP)
107 (ÜP)
Probe insulation
Device or probe defective.
Contact the manufacturer and specify the trigger for diagnostic purposes.
Help:
- Remove the probe and check it in ambient air.
- Check the wiring between LT3-F and probe (sensor signal
and terminals 10/11/12)
- Probe and LT3-F at the same potential?
- Exchange the probe.
008 (HP)
108 (ÜP)
Intern fault: measured value comparison HP/UP
Device defective; contact the manufacturer and specify the trigger for diag-
nostic purposes.
014 (HP)
114 (ÜP)
Write the data record
015 (HP)
115 (ÜP)
Parameter/EEPROM fault
Device defective. Contact the manufacturer and specify the trigger for
diagnostic purposes.
016 (HP)
116 (HP)
Internal fault/self tests
Device defective. Contact the manufacturer and specify the trigger for
diagnostic purposes.
Содержание LT3-F
Страница 2: ......
Страница 14: ...13 3 Product Description 3 2 Declaration of Conformity...
Страница 15: ...14 3 Product Description...
Страница 20: ...19 5 Display and Operational Controls 5 2 2 Information Menu Structure 5 2 3 Calibration Menu Structure...
Страница 21: ...20 5 Display and Operational Controls 5 2 4 Settings Menu Structure...
Страница 24: ...23 5 Display and Operational Controls 5 4 1 Main Menu Password Entry...