Press
Q
to select
CONTINUE
and the calibration
will be calculated for future recall from the
library.
51. H2O
DONE HI STD
*CONTINUE
SCAN AGAIN
The
EDIT USER TEST 3
menu will appear.
EDIT USER TEST 3
*FORMAT RESULT
SELECTING THE NUMERICAL FORMAT OF THE RESULT
To input tests with very different ranges, the number of decimal places
displayed for a result can be selected. A test which ranges from 20 to
1000 ppm should not be displayed with three decimal places. A test with
a range from 0.010 to 0.500 needs three decimal places (the
microprocessor will always calculate the concentration to many more
significant figures than will be displayed). Menu choices of 0, 1, 2, or 3
decimal places will be given for the display.
Press
Q
to select
FORMAT RESULTS
.
EDIT USER TEST 3
*FORMAT RESULT
44