background image

3

DC1200 - INSTRUMENT PERFORMANCE CHECK 

Perform this procedure at specifi ed intervals to verify instrument performance.

NOTE: Store standards in a cool place out of direct light. Exposure to excessive heat 
or light may cause degradation of the standards.

Wipe the DPD Chlorine Color Standard 
Blank (4136-E) with a clean, lint-free cloth.

Insert the DPD 
Chlorine Color 
Standard Blank into 
the chamber, being 
sure to align the 
index line with the 
arrow on the meter. 
Close the lid.

Push the 

READ

 

button and hold it 
for 2 seconds until 

bLA

 is displayed. 

Release the button 
to take a zero 
reading.

Wipe the DPD Chlorine Color Standard 
A (4147-E) with a clean, lint-free cloth.

Press the 

READ

 

button. The 
concentration in 
ppm will be 
displayed within 
2 seconds. The 
reading should be 
within the tolerance 
limits given on the 
Data Sheet.

Repeat steps 4-6 with DPD 
Chlorine Color Standard B 
(4138-E) and DPD Chlorine 
Color Standard C (4139-E).

If the readings are not within the 
tolerance limits, standardize the 
meter calibration by following 
the Calibration Procedure in the 
meter manual. If the readings 
are still not within the tolerance 
limits, the standards are no longer 
valid and should be replaced.

Insert the DPD 
Chlorine Color 
Standard Blank into 
the chamber, being 
sure to align the 
index line with the 
arrow on the meter. 
Close the lid.

1

2

3

4

5

6

7

8

Содержание 4140-03

Страница 1: ...e to be used as Chlorine Secondary Standards to verify the performance of the meter and are not intended to be used as calibration standards The meter is calibrated at the time of manufacture The stan...

Страница 2: ...loth Press the READ button The concentration in ppm will be displayed within 2 seconds Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Push the READ button to turn the m...

Страница 3: ...e a zero reading Wipe the DPD Chlorine Color Standard A 4147 E with a clean lint free cloth Press the READ button The concentration in ppm will be displayed within 2 seconds The reading should be with...

Страница 4: ...he concentration will be displayed 015 Chlorine Scan Sample Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press ON to turn the meter on Press ENTER to select START Pre...

Страница 5: ...o the chamber being sure to align the index line with the arrow on the meter Close the lid Repeat steps 4 6 with DPD Chlorine Color Standard B 4138 E and DPD Chlorine Color Standard C 4139 E If the re...

Страница 6: ...ill be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Menu Press...

Страница 7: ...he readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within the tolerance limits t...

Страница 8: ...ration will be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Me...

Страница 9: ...r Close the lid If the readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within th...

Страница 10: ...the dye standards being slightly different from the spectral qualities of DPD Readings of actual DPD chlorine reactions will not differ significantly between meters Contact LaMotte Company Technical...

Страница 11: ...BLANK...

Страница 12: ...TE COMPANY Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800 344 3100 410 778 3100 Outside U S A Fax 410 778 6394 Visit us on the web at www lamotte com 64140 03...

Отзывы: