Содержание 4140-03

Страница 1: ...e to be used as Chlorine Secondary Standards to verify the performance of the meter and are not intended to be used as calibration standards The meter is calibrated at the time of manufacture The stan...

Страница 2: ...loth Press the READ button The concentration in ppm will be displayed within 2 seconds Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Push the READ button to turn the m...

Страница 3: ...e a zero reading Wipe the DPD Chlorine Color Standard A 4147 E with a clean lint free cloth Press the READ button The concentration in ppm will be displayed within 2 seconds The reading should be with...

Страница 4: ...he concentration will be displayed 015 Chlorine Scan Sample Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press ON to turn the meter on Press ENTER to select START Pre...

Страница 5: ...o the chamber being sure to align the index line with the arrow on the meter Close the lid Repeat steps 4 6 with DPD Chlorine Color Standard B 4138 E and DPD Chlorine Color Standard C 4139 E If the re...

Страница 6: ...ill be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Menu Press...

Страница 7: ...he readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within the tolerance limits t...

Страница 8: ...ration will be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Me...

Страница 9: ...r Close the lid If the readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within th...

Страница 10: ...the dye standards being slightly different from the spectral qualities of DPD Readings of actual DPD chlorine reactions will not differ significantly between meters Contact LaMotte Company Technical...

Страница 11: ...BLANK...

Страница 12: ...TE COMPANY Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800 344 3100 410 778 3100 Outside U S A Fax 410 778 6394 Visit us on the web at www lamotte com 64140 03...