2020 ClearTrace

In-line Turbidity System

Code 5-0113-24V

Code 5-0113-110V

Содержание 2020 ClearTrace Series

Страница 1: ...2020 ClearTrace In line Turbidity System Code 5 0113 24V Code 5 0113 110V...

Страница 2: ......

Страница 3: ...r concerns The information presented in this instruction manual is subject to change without notice as improvements are made and does not represent any commitment whatsoever on the part of LaMotte LaM...

Страница 4: ...ed or authorized person should carry out installation setup and operation of the system Before using the system make sure that is connected as specified Failure to do so may result in permanent damage...

Страница 5: ...cement Parts 8 3 Mounting Wiring and Plumbing 9 3 1 Mounting 9 3 2 Plumbing 10 3 3 Valve Position for Operation 10 3 4 Wiring For 24 VDC and 110VAC Power 11 3 5 Sample Chamber Details 14 4 Measure Scr...

Страница 6: ...and specifications must be read and understood by all persons involved in installation and operation of this equipment This manual identifies safety instructions and additional information by means of...

Страница 7: ...e AV38 electronics and cabling Observing the prescribed environmental and operational conditions Observing chemical compatibility with all wetted materials 1 3 Safety Instructions The Turbidity System...

Страница 8: ...volume is only 135 milliliters only a small amount of a turbidity standard is necessary to perform EPA mandated calibrations A white light source in the chamber lid illuminates suspended particles wit...

Страница 9: ...be Code 5 0116 Turbidity standard 0 0 NTU 1 liter Code 5 0117 Turbidity standard 0 0 NTU 1 gallon Code 5 0118 Turbidity standard 1 0 NTU 1 liter Code 5 0119 Turbidity standard 1 0 NTU 1 gallon Code 5...

Страница 10: ...hen handled or touched Mounting tools and equipment inch Drill with drill bit for appropriate wall anchor Four diameter bolts 1 5 inches long Four inch ID flat washers Wrench to turn bolts Level to in...

Страница 11: ...at result in back pressure for OUTLET and DRAIN are not allowed INLET Plumbing Recommendations Use 3 8 ID tube or inch NPT threaded pipe from sample source Flow rate must be between 250 to 750 ml min...

Страница 12: ...power and typically draws up to 250 mA 10 watts Pg11 and Pg16 punch out conduit holes are provided for user terminal connections We supply Pg11 cord grips that can be removed Note Loop 2 and Relay op...

Страница 13: ...ace Tubidimeter comes with a Junction Box below the two terminal blocks when ordered with one or two 4 20 mA analog outputs and or host comminications If relays are ordered a third terminal block is p...

Страница 14: ...ire from Host Network Network Ground 3 White wire from Host Network 4 Blue wire from Host Network 5 Shield wire from Host Network 6 OUTPUT 1 7 OUTPUT 1 8 OUTPUT 2 9 OUTPUT 2 10 Relay Normally Open Ter...

Страница 15: ...cientific Inc and its subsidiaries 14 Exterior marker to align lid with sample chamber Sample Chamber Lid with integral lamp assembly Alignment hole The alignment peg on the Sample Chamber must line u...

Страница 16: ...cations adapter are connected to the AV38 display but the optical sensor head is absent the measure screen indicates the fault as shown in Figure 3 2 When the DataStick or communications adapter are n...

Страница 17: ...t 4 mA and 20 mA values Calibrate and test the current output with a meter Analog Output 2 option Not currently available PID Controller Not currently available Relay A option Not currently available...

Страница 18: ...turn to the measure screen Pressing the down arrow from Configure will display DataStick Comms To select DataStick Comms functions press ENTER To continue scrolling the list press the down arrow Press...

Страница 19: ...t AV38 To select Reset functions press ENTER To continue scrolling the list press the down arrow Press ESC to return to the measure screen To return to the measure screen press ENTER when in the EXIT...

Страница 20: ...ternative a formazin 20 NTU calibration standard may be prepared from a stock solution of 4000 NTU formazin and used within 24 hours Refrigerate when stored 7 After filling the sample chamber with the...

Страница 21: ...hen ready press the ENTER key The calibration value can be adjusted with the arrow keys Use the up down arrows for fine adjust and the left right arrows for course adjust After the calibration value i...

Страница 22: ...reen If the calibration is not successful pressing ENTER will return to the beginning of the 1 point sample calibration Reasons for calibration failure include a sample that is not in the correct meas...

Страница 23: ...nd place it on top of the AV38 4 Rinse the chamber with clean water and then with a small amount of calibration solution See calibration kit instructions Code 5 0114 or code 5 0015 for procedure 5 Shu...

Страница 24: ...alibration is okay the Measure screen will be displayed NOTE 0 000 NTU may not be displayed An acceptable reading could be as high as 0 015 NTU If an error occurs during calibration the reason for the...

Страница 25: ...o calibrate temperature when the sample chamber is full of flowing process water When the value has been edited as desired the enter key is pressed and the calibration of the point is started During t...

Страница 26: ...rocedure If an error occurs during calibration that causes the procedure to fail the reason for the failure will be shown When the enter key is pressed the calibrate menu is displayed This ends the te...

Страница 27: ...with the up down arrows Press ENTER to select a new filter value Press ESC to abort the new sensor value The temperature filter edit screen works the same way The Temperature units selection offers C...

Страница 28: ...or precise current readings When the AV38 is configured with two current outputs then the main menu list shows Analog Output 1 and Analog Output 2 The menu system for each output is identical Note Dur...

Страница 29: ...the displayed value with the ENTER key Exit without changing by using the ESC key To set the 20 milliamp value select Set 20 mA Value with the up down arrows and the ENTER key Use the up down arrows f...

Страница 30: ...he displayed count value with the up down arrows until the meter reads exactly 4 000 mA Accept the displayed value with the ENTER key Exit without changing by using the ESC key To calibrate the 20 mA...

Страница 31: ...d the firmware version To view the current software version of the AV38 press ENTER at each screen level as shown in the diagrams below 30 Main Menu Help MENU ESC ENTER t t t t t Help About MENU ESC E...

Страница 32: ...sure the user acknowledges that a reset has been done 13 PERFORMANCE SPECIFICATION Measurement Range 0 to 100 NTU Can range to 200 NTU when necessary Resolution 0 001 NTU Accuracy 2 of reading or 0 0...

Страница 33: ...ml Light Source Housing Material Anodized Aluminum Mounting Plate 12x12 inches 4 mounting holes Sensor Head Material Quartz Glass Anodized Aluminum Weight 5 6 lbs Units of Measure Measurement Units N...

Страница 34: ...eglect lack of appropriate maintenance alteration accident or improper application or installation 2 Damage caused by any repair or attempted repair not authorized by LaMotte 3 Any product not used in...

Страница 35: ......