13
During the sterilization cycle, the user can press the button Start/Stop and keep 5 seconds to stop the whole cycle.
5. Installation
The autoclave is packed by a wooden carton. To open the carton, you need a flat screwdriver to open the cover. Then, the
autoclave can be lifted out by 2 persons.
5.1) When installing the autoclave onto the table, please keep the autoclave at least 10 cm away from surroundings. And
keep the rear side of the autoclave at least 20cm away from the surrounding.
It is recommended to install the autoclave in a well-
ventilated location. Don’t block the radiator of the st
erilizer.
Keep the table stable & horizontal.
5.2) Adjust atmospheric pressure
Because different countries have different atmospheric pressure, the user shall adjust the atmospheric pressure before
first use.
Adjust method:
Cut off power
Open the door of autoclave
Turn on the power, and keep it for 20 seconds. Then cut it off.
Atmospheric pressure for autoclave is successfully adjusted.
5.3) Adjust time and date
Please refer to information stated in 4.3.7.
Please ensure the table for installation is strong enough for an autoclave.
The autoclave might not run if the user does not adjust its atmospheric pressure.