8
4.2 Introduction of
“
USER MENU
”
There are 11 sterilization programs for user to choose.
➢
121
℃
-program and 134
℃
-program has no difference in sterilization results. If the instruments cannot be sterilized in
134
℃
, please choose 121
℃
program.
➢
Solid program can only sterilize solid instruments without wrapped, such as pliers, forfex, forceps etc.
➢
Porous program can sterilize the loads which made of porous material
➢
Hollow program can sterilize the hollow A and hollow B loads
➢
User-defined program: All this program's parameters can be adjusted: The sterilization temperature
、
the sterilization
time
、
dry time and vacuum times.
➢
B&D test is for hollow A loads test . B-D test and Helix test is the same test program. The only difference is that B-D
test uses a B-D test package, while Helix test uses a Helix test device (PCD:process challenge device).
➢
Vacuum test is air leakage test.
➢
Clean program: This is used for clean the autoclave. If the autoclave shows
“
NEED CLEAN
”
, please run this
program. This program can also be used to clean other loads.
➢
Prion program: This program is to sterilize prions.
4.3 ADV MENU
It is recommended to do a vacuum test every mouth. If the result is FAIL, Do not use
this autoclave.