No.:GC-JS-21 Version: 01/00
Effective date
:
2020-07-22
17
5. Introduction of preset programs
5.1 Parameters of preset programs
Program
Sterilization
temperature
(
℃
)
)
Holding time
(
Min
)
Pre-vacuum
times
Drying time
(
Min
)
Remarks
134℃
universal quick
134
4
3
4
Preset
134℃
universal
134
4
3
6
Preset
134℃ solid
134
4
1
2
121℃
universal
121
4
3
8
Preset
B-D/Helix test
134
3.5
3
4
Test program
Vacuum test
Vacuum to -80kPa
,
holding 15min
Test program
Drying
5min
Preset drying
5.2 Introduction to the load of each program
5.2.1 Load type corresponding to each program
134℃
universal
quick
Non-wrapped load or wrapped solid, fabric and cavity load, Paper-plastic packaging
or double-layer packaging may not dry, should be used immediately after sterilization
Universal
Non-wrapped load or wrapped solid, fabric and cavity load
134℃ solid
Solid load without package, should be used immediately after sterilization
B-D/Helix test
B-D test device or PCD test device
Vacuum test
Without any loading
Drying
Dry the load when it is not dry after sterilization
5.2.2 Applicable maximum load table
Model
Maximum weight of instruments
Maximum weight of fabric loads
LFSS03AA
2.2kg
0.5kg /package
5.3 Maximum working time and maximum water consumption when each program is at
maximum load
Model
Loading
kg
134℃ Universal
quick
134℃ Universal
121℃ Universal
134℃ Solid
Longes
t time
Min
water
consump
Longest
time
Min
water
consump
Longest
time
Min
water
consump
Longest
time
Min
water
consump