Range:
The MMT features a dual measurement
range. This allows the MMT to have better
resolution for measurements of smaller
muscle groups.
Range Specifications:
Low Range:
0-50lbs. with a resolution of 0.1lbs.
0-22.6kg with a resolution of 0.1kg
High Range:
0-300lbs. with a resolution of 0.4lbs.
0-136.1kg with a resolution of 0.2kg
The range is changed by pressing the RANGE button. An “H” or “L” indicator on the Main
Measurement Screen shows the range setting for the MMT. The MMT is automatically reset
when the range is changed.
Caution 1:
Caution 2:
When using the low range setting, any reading over 50lbs. (22.6kg) is outside the
range and should be disregarded as an erroneous reading.
Do not exceed the 300 pound force limit. Excessive force may cause permanent
damage to the MMT or alter the calibration.
Caution:
Since the RESET button re-zeros the MMT, force should not be applied while the
RESET button is pressed.
Reset:
The Reset function clears the display screens
and sets the MMT for a new measurement.
A reset is also required whenever a
Diagnostic Error message is displayed. A
reset can be triggered by several different
functions.
Reset Guidelines:
1. The MMT is automatically reset when the power is applied.
2.
A reset is also required whenever a Diagnostic Alert Message is displayed.
3. The MMT is automatically reset when the range is changed.
4. The MMT is automatically reset after the Calibration Routine is completed.
5. The Battery Saver Mode does not cause a Reset to occur. Data is preserved after the
Battery Saver Mode is terminated.
6. Running the Main Menu does not cause a Reset to occur.
7. The MMT automatically resets on the application of a new force.
8. Pressing the RESET button will clear the display and set the MMT for a new
measurement.
9. Pressing the RESET button also sets the zero point for measurements.