Troubleshooting
27
Symptom
Error codes
Possible solution
Critical technical
error. Error code
is shown in the
display and error
light activated.
x01 - Program memory
x02 - Calibration memory
x04 - Data memory
x08 - RTC crystal
x10 - Display driver
communication
x20 - Accelerometer
communication
x40 - Light sensor
communication
x80 - ECG analog signal
chain
Continue standard
procedure without use of
the device.
Use alternative means of
measuring newborn heart
rate, e.g. a stethoscope.
Place the device on the
charging stand. If the
problem persists, remove
device from service.