22
Reprocessing
Clean and disinfect NeoBeat after each patient use to minimize the risk of
cross-contamination.
Warning
Do not place a used NeoBeat back onto the charging stand before it has been
cleaned and disinfected.
Cleaning
1. Clean all surfaces of NeoBeat using a cloth dampened with soapy water.
To remove difficult soil, use a toothbrush dipped in lukewarm soapy
water. Clean for a minimum of 2 minutes, ensuring that all soil has
been removed.
2. Wipe NeoBeat with a clean cloth dampened with clean water.
3. Dry NeoBeat using a clean cloth or by air drying.
Disinfection
1. Wipe all surfaces of NeoBeat with a clean cloth soaked with 70%
ethanol.
2. Spray 70% ethanol on all surfaces of NeoBeat. Ensure it remains
wet for 10 minutes. Repeat spraying as necessary to account for
evaporation.
3. Wipe NeoBeat with a clean cloth dampened with clean water.
4. Allow to dry completely.