Troubleshooting
26
Symptom
Possible cause
Possible solution
Patient contact is
not detected.
The skin is too dry for NeoBeat to
detect patient contact.
Wet the electrodes
with water and/or
reposition the device.
Heart rate is not
detected.
The device is misplaced or not in
direct contact with patient.
Or
Stimulation/movement/
handling of patient temporarily
generating too much noise.
Or
Heart rate is below 20 bpm.
Reposition the
device. Moving the
device closer to the
heart may give a
stronger signal.
If the problem
continues, use
alternative means of
measuring newborn
heart rate, e.g. a
stethoscope.