background image

 

User Manual LabModul 240

 

Page 3

 

www.labtec.de

 

3.  device description

 

 

LabModul  240 

are  electronic  test  devices  to 

be  operated  via  a  PC  for  the  testing  of  brea-

thing  masks  according  to  DIN  EN  136,  self

-

contained  breathing  apparatus  (SCBA)  ac-

cording to DIN EN 137 and chemical protecti-

ve suits (CPS) according to DIN EN 943

-

1.

 

 

The  tests  are  performed  in  the  low  (

-

50  to  50 

mbar) and medium pressure measuring range  

(0 to 20 bar).

 

 

For  testing  other  breathing  apparatus  or  che-

mical  protective  suits,  contact  Labtec.  Only 

breathing air according to DIN EN 12021  may 

be used.

 

 

Module  T

 Control panel with LCD display and 

membrane  keyboard  for  manual  operation  of 

the device.

 

 

Module  COM 

Freely positionable touchscreen 

PC,  connected  to  test  device  via  swivel  arm 

mounting,  incl.  pre

-

installed  LabNet  Profi  test 

software

 

 

The device is also available as model MT. The 

model 

LabModul  240  MT 

has no interface to 

the LabNet Profi operating software but is de-

signed for purely manual operation without PC 

support. The LabModul 240 MT has the stand-

ard operating  panel  of option T. The interface 

can be retrofitted at the factory. 

 

 

Connection requirements

 

 

A 230

-

240 V / 50

-

60 Hz power supply is requi-

red to operate the test stand.

 

 

As medium pressure connection Euro

-

coupling 

and  nipple  are  available  as  standard.  If  units 

with other coupling connections are tested, ap-

propriate intermediate adapters must be used. 

Please  contact  the  manufacturer  of  the  units 

for this. 

 

 

 

Scope of delivery and accessories 

 

 

The scope of delivery of the device includes a 

power supply cable and a USB cable.

 

 

Adapters for the connection of CPS, regulators 

and  sealing  caps  for  respirators  must  be  or-

dered separately depending on the connection 

used  (e.g.  ESA,  PE45,  RD40,  manufacturer

-

specific connections).

 

 

 

 

Labtec  offers  a  wide  range  of  accessories  for 

test devices.

 

 

The range  includes test  adapters for adapting 

all  respiratory  protection  devices  on  the  mar-

ket,  quick  release  devices  for  quick  device 

adaptation  as  well  as  holders  for  holding  hel-

met

-

mask combinations

 

4. technical data

 

Medium pres-

sure connec-

tion:

 

EURO coupling and nipple 0...20 bar  

 

Medium pres-

sure sensor:

 

Measuring range

 

precision

 

0...25 bar

 

± 0,5 %

 

Low pressure 

sensor:

 

Measuring range

 

precision

 

-

50.0.+50 mbar

 

± 0,05 %

 

Dead space

 

approx. 480 ml 

-

 w. o. Adapter 

 

Power 

 

Transformer

 

Input:         230 V AC

 

Output:        24 V DC

 

Dimensions

 

~ B 460 x H 490 x D 430 mm 

 

Weight:

 

approx. 21 kg

 

 

Содержание LabModul 240 Com

Страница 1: ...2 3 device description 3 4 technical data 3 5 operating elements 4 6 component description 5 7 description of the operating elements 5 8 commissioning 6 9 care and maintenance 6 Attachments Performin...

Страница 2: ...ed and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees training partners and service technicians i...

Страница 3: ...fitted at the factory Connection requirements A 230 240 V 50 60 Hz power supply is requi red to operate the test stand As medium pressure connection Euro coupling and nipple are available as standard...

Страница 4: ...10 l min switching valve 9 Inhalation 10 l min switching valve 10 Fill CPS switching valve 11 Empty CPS switching valve 12 Relieve medium pressure tactile valve A test stand housing B Panel optional C...

Страница 5: ...um pump With the built in vacuum pump the required air currents are generated and the probe is infla ted or deflated The desired air flow rate can be set in the software from 2 12 l min Processor The...

Страница 6: ...g nipple of the medium pressure device connection and the line of the demand valve to the EURO quick coupling of the medium pres sure device connection If the SCBA to be tested does not have a Euro co...

Страница 7: ...specific adaptors 1 2 2 Mount mask on test head pull straps tight inflate test head 3 A few seconds is usually long enough 1 2 3 Adjust negative pressure to exactly mbar 5 By means of pressure relief...

Страница 8: ...with the setpoint All further tests require high pressure Regulator in conjunction with a Connect the SCBA to the test device Medium pressure hose from the PA regulator of a SCBA must with Device conn...

Страница 9: ...ssure flows into the device 3 2 2 Close the bottle of the SCBA Tester DUT are under pressure 3 2 3 Read the medium pressure on the LCD display MP and compare with setpoint 3 2 4 After a specified time...

Страница 10: ...pressure is reached 5 1 8 Lower the pressure to the required internal pres sure with the switch 11 if necessary with but ton 12 approach and turn on the stopwatch Setting the required Settling time 5...

Страница 11: ...h in EEC Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l...

Отзывы: