Page 9

 

www.labtec.de

 

3

 

Pressure reducer

 

3.1

 

Medium pressure, static at #

 

bar inlet pressure

 

3.2

 

Medium pressure, dynamic at # l/min ventilation

 

3.3

 

Medium pressure, dynamic at 10 l / min extraction (optional, not required by manufacturers) 

 

 

Test 

step

 

Operation

 

Comments

 

Controls

 

3.1.1

 

Open the bottle of the SCBA .

 

Medium pressure flows into the device.

 

 

3.1.2

 

Close the bottle of the SCBA .

 

Tester / DUT are under pressure.

 

 

3.1.3

 

Read the medium pressure on the LCD display 

(MP) and compare with setpoint. 

 

 

 

 

 

 

 

3.2.1

 

Open the bottle of the SCBA .

 

Medium pressure flows into the device.

 

 

3.2.2

 

Close the bottle of the SCBA .

 

Tester / DUT are under pressure.

 

 

3.2.3

 

Read the medium pressure on the LCD display 

(MP) and compare with setpoint. 

 

 

 

3.2.4

 

After a specified time of # min, read the medium 

pressure again and compare with the previously 

measured value. 

 

Compare difference value with allo-

wable deviation. 

 

 

 

 

 

 

3.3.1

 

Open the bottle of the SCBA and leave it open. 

 

Medium pressure flows into the device 

and is in the device. 

 

 

3.3.2

 

Switch on pump with switch (9) and flow of 

-

10 l / 

min. produce. 

 

Regulator must be connected. 

 

 

3.3.3

 

Read medium pressure on the LCD (MP) dis-

play. 

 

 

 

3.3.4

 

Close the bottle and relieve the medium pres-

sure with the button (12). 

 

at the end of the test or continue with 

further tests. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A 3

 

testing breathing apperatus

 

Содержание LabModul 240 Com

Страница 1: ...2 3 device description 3 4 technical data 3 5 operating elements 4 6 component description 5 7 description of the operating elements 5 8 commissioning 6 9 care and maintenance 6 Attachments Performin...

Страница 2: ...ed and certified Labtec service technician and tested and calibrated by him or her using a Labtec calibration device The list of current certified employees training partners and service technicians i...

Страница 3: ...fitted at the factory Connection requirements A 230 240 V 50 60 Hz power supply is requi red to operate the test stand As medium pressure connection Euro coupling and nipple are available as standard...

Страница 4: ...10 l min switching valve 9 Inhalation 10 l min switching valve 10 Fill CPS switching valve 11 Empty CPS switching valve 12 Relieve medium pressure tactile valve A test stand housing B Panel optional C...

Страница 5: ...um pump With the built in vacuum pump the required air currents are generated and the probe is infla ted or deflated The desired air flow rate can be set in the software from 2 12 l min Processor The...

Страница 6: ...g nipple of the medium pressure device connection and the line of the demand valve to the EURO quick coupling of the medium pres sure device connection If the SCBA to be tested does not have a Euro co...

Страница 7: ...specific adaptors 1 2 2 Mount mask on test head pull straps tight inflate test head 3 A few seconds is usually long enough 1 2 3 Adjust negative pressure to exactly mbar 5 By means of pressure relief...

Страница 8: ...with the setpoint All further tests require high pressure Regulator in conjunction with a Connect the SCBA to the test device Medium pressure hose from the PA regulator of a SCBA must with Device conn...

Страница 9: ...ssure flows into the device 3 2 2 Close the bottle of the SCBA Tester DUT are under pressure 3 2 3 Read the medium pressure on the LCD display MP and compare with setpoint 3 2 4 After a specified time...

Страница 10: ...pressure is reached 5 1 8 Lower the pressure to the required internal pres sure with the switch 11 if necessary with but ton 12 approach and turn on the stopwatch Setting the required Settling time 5...

Страница 11: ...h in EEC Guideline 2016 42 EG and EMC 2014 30 EU including all changes and addendums to date thereto The above mentioned product meets the following standards and guidelines Nous Vous Confirmons que l...

Отзывы: