SX-01635-000, Rev. 5
26
Appendix A Specifications
Radiometer Performance
Temperature Coefficient:
+(0.02%of r 0.5 counts) per
°
C
Input Bias Current:
0.2 pA maximum
Current Input Characteristics at 25°C
Range
Resolution
Accuracy (% r counts)
24 hr period
Accuracy (% r counts)
1 yr period
1 nA
0.1 pA
0.05 + 2
0.1 + 5
10 nA
1 pA
0.05 + 1
0.1 + 2
100 nA
10 pA
0.05 + 1
0.1 + 2
1 µA
100 pA
0.05 + 1
0.1 + 2
10 µA
1 nA
0.05 + 1
0.1 + 2
100 µA
10 nA
0.05 + 1
0.1 + 2
1 mA
100 nA
0.05 + 1
0.1 + 2
Voltage Input Characteristics at 25°C
Range
Resolution
Accuracy (% r counts)
24 hr period
Accuracy (% r counts)
1 yr period
10 mV
1 µV
0.05 + 2
0.1 + 5
100 mV
10 µ V
0.05 + 1
0.1 + 2
1 V
100 µV
0.05 + 1
0.1 + 2
10 V
1 mV
0.05 + 1
0.1 + 2
50 V
5 mV
0.05 + 1
0.1 + 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com