SX-01635-000, Rev. 5
19
Calibrations
C
n
C
n
X
Typing
C
and the number that corresponds to a calibration stored in the SC-5500 sets the
instrument for that specific calibration. (
n
) is the one or two digit decimal number of the
calibration. Example:
C0X
sets the SC-5500 to Cal 0,
C1X
sets the SC-5500 to Cal 1.
(C0 is factory set to display
volts
, C1 is factory set to display
amps
.)
Data Relay
O
n
OØX
Tells the SC-5500 to relay only one detector measurement to the computer for display.
(This is not allowed for IEEE 488 operation, but is required for RS-232 operation.)
O1X
Tells the SC-5500 to relay continuous detector measurements for display. (This is not
allowed for IEEE 488 operation, but is required for RS-232 operation.)
Digit Display
H
n
HØX
Formats the detector readings for 3-1/2 digits on display and over computer interface.
H1X
Formats the detector readings for 4-1/2 digits on display and over computer interface.
Digital Filter
F0X
Enables digital filter
F1X
Disables digital filter
Digital Output Control P
n
J
m
(K
m
)
Pn J
m
X Sends a +5 volt charge to the pin that corresponds with the bit (
J
m
) of the port (
P
n
). (
n
)
is a decimal number from 1 to 4; (
m
) is a decimal number from 1 to 8.
Example:
P2 J3X
Sets bit 3 of port 2 high
P
n
K
n
XShuts down the voltage to the pin that corresponds with the bit (
K
n
) of the port (
P
n
).
Example:
P2 K3X
Sets bit 3 of port 2 low.
Normalizing Data
N
n
NØX
Sets the SC-5500 in the normalize mode.
N1X
Turns off the normalize mode.
Manual Ranging
R
n
R
n
X
Sets the dynamic range of the SC-5500's detector measurements. When measuring cur-